ROTATIONAL RAMAN CALIBRATION OF THOMSON SCATTERING

被引:42
作者
HOWARD, J
JAMES, BW
SMITH, WIB
机构
[1] School of Phys., Univ. of Sydne, Sydney, NSW
关键词
D O I
10.1088/0022-3727/12/9/005
中图分类号
O59 [应用物理学];
学科分类号
摘要
Anti-Stokes rotational Raman scattering from nitrogen gas at near atmospheric pressure and room temperature has been used to calibrate the intensity of ruby laser light Thomson scattered from a plasma in a transverse magnetohydrodynamic shock tube. The main limit to the accuracy of this calibration method is the uncertainty in the available Raman cross section for nitrogen.
引用
收藏
页码:1435 / 1440
页数:6