COMPARISON OF POTENTIOMETRIC ENZYME SENSORS FOR UREA AND PENICILLIN-G - DIFFERENTIAL THICK-FILM SENSORS VERSUS CLASSICAL ELECTRODES

被引:12
作者
EPPELSHEIM, C
AUBECK, R
HAMPP, N
机构
[1] Institute for Physical Chemistry, University of Munich, D-80333 München
关键词
UREASE; PENICILLINASE; BIOSENSOR; THICK-FILM; ELECTRODE;
D O I
10.1016/0376-7388(94)00256-X
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
A comparative study of urease and penicillinase biosensors was performed in which different transducers were compared, i.e. pH glass electrodes, macroscopic ion-selective electrodes and thick-film transducers. For enzyme immobilization onto the transducers aqueous dispersion polymers were used in all cases in this work. Since these sensors are based on potentiometric measurements the lack of a classical reference electrode when solid state transducers are employed is a crucial feature. Results obtained in this work revealed that thick-film sensors without a classical reference electrode can compensate this deficit when they operate in a differential mode. This means that it is not the absolute potential change of a single selective electrode, induced by the analyte concentration, that is measured but rather the relative change with reference to a non-selective electrode. Enzyme sensors with macroscopic transducers and classical reference electrodes showed the same basic characteristics compared with thick-film sensors. The long term stability of the sensors is sufficient; urease and penicillinase thick-film sensors showed 50 and 75% of the initial response after two weeks of permanent use at room temperature.
引用
收藏
页码:131 / 137
页数:7
相关论文
共 14 条
  • [1] ANZAI J, 1989, MAKROMOL CHEM-RAPID, V10, P167
  • [2] ANZAI J, 1987, CHEM PHARM BULL, V35, P4568
  • [3] POTENTIOMETRIC THICK-FILM SENSOR FOR THE DETERMINATION OF THE TUMOR-MARKER BOUND SIALIC-ACID
    AUBECK, R
    EPPELSHEIM, C
    BRAUCHLE, C
    HAMPP, N
    [J]. ANALYST, 1993, 118 (11) : 1389 - 1392
  • [4] BUSH K, 1984, METHOD ENZYMAT AN, V2, P280
  • [5] POTENTIOMETRIC ION-SELECTIVE AND BIO-SELECTIVE ELECTRODES BASED ON ASYMMETRIC CELLULOSE-ACETATE MEMBRANES
    CHA, GS
    MEYERHOFF, ME
    [J]. TALANTA, 1989, 36 (1-2) : 271 - 278
  • [6] ION-SELECTIVE ELECTRODES FOR THE DETERMINATION OF THE ANTIARRHTHYMIC DRUG BRETYLIUM
    EPPELSHEIM, C
    BRAUCHLE, C
    HAMPP, N
    [J]. ANALYST, 1992, 117 (10) : 1609 - 1612
  • [7] DESIGN AND APPLICATION OF THICK-FILM MULTISENSORS
    HAMPP, N
    EPPELSHEIM, C
    POPP, J
    BISENBERGER, M
    BRAUCHLE, C
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 1992, 31 (1-3) : 144 - 148
  • [8] HAMPP N, 1992, GBF MONOG SERIES, V17, P153
  • [9] IUPAC Analytical Chemistry Division, 1976, PURE APPL CHEM, V48, P127, DOI [DOI 10.1351/PAC197648010127, 10.1351/pac197648010127]
  • [10] Kerscher L, 1985, METHOD ENZYMAT AN, V8, P444