IMPEDANCE BASED SENSING OF THE SPECIFIC BINDING REACTION BETWEEN STAPHYLOCOCCUS ENTEROTOXIN-B AND ITS ANTIBODY ON AN ULTRA-THIN PLATINUM FILM

被引:65
作者
DESILVA, MS
ZHANG, Y
HESKETH, PJ
MACLAY, GJ
GENDEL, SM
STETTER, JR
机构
[1] UNIV CHICAGO,DEPT ELECT ENGN & COMP SCI,MICROFABRICAT APPLICAT LAB,CHICAGO,IL 60607
[2] US FDA,NATL CTR FOOD SAFETY & TECHNOL,SUMMIT ARGO,IL 60501
关键词
BIOSENSOR; IMMUNOBIOSENSOR; STAPHYLOCOCCUS ENTEROTOXIN B; PLATINUM ULTRA-THIN FILM;
D O I
10.1016/0956-5663(95)96958-2
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Immunobiosensing techniques to measure specific antigen-antibody binding reactions are important in the development of biosensor applications in biotechnology, in vitro diagnosis, medicine and food technology. An immunobiosensor was constructed to measure the specific binding reaction between Staphylococcus enterotoxin B (SEE) and anti-SEE antibodies. The biosensor comprised an anti-SEE bioactive layer covalently immobilized on an ultra-thin platinum (Pt) film sputtered onto a 100 nm thick silicon dioxide layer on a silicon chip. The Pt film was discontinuous with a normal thickness of 25 Angstrom. The impedance of the Pt film decreased during the binding of the anti-SEE to SEE in phosphate buffered saline (PBS) at room temperature. The impedance decreases were irreversible in PBS before saturation of the specific binding sites. When saturated, the impedance at 100 Hz was 14% of the value obtained for the fresh anti-SEE layer in PES. The magnitude of the impedance (/Z/) decrease followed a simple relationship with SEE concentration in the range between 0.389 and 10.70 ng/ml SEE. The specificity of the biosensor was demonstrated by showing that no irreversible impedance decreases occurred when the sensor was exposed to 100 ng/ml kappa-casein, alpha-casein, or alpha-lactalbumin, in PBS.
引用
收藏
页码:675 / 682
页数:8
相关论文
共 15 条
  • [1] AIZAWA M, 1986, 2ND P INT M CHEM SEN, V6, P622
  • [2] DIRECT DETECTION OF IMMUNOSPECIES BY CAPACITANCE MEASUREMENTS
    BATAILLARD, P
    GARDIES, F
    JAFFREZICRENAULT, N
    MARTELET, C
    COLIN, B
    MANDRAND, B
    [J]. ANALYTICAL CHEMISTRY, 1988, 60 (21) : 2374 - 2379
  • [3] USE OF THIOL-TERMINAL SILANES AND HETEROBIFUNCTIONAL CROSSLINKERS FOR IMMOBILIZATION OF ANTIBODIES ON SILICA SURFACES
    BHATIA, SK
    SHRIVERLAKE, LC
    PRIOR, KJ
    GEORGER, JH
    CALVERT, JM
    BREDEHORST, R
    LIGLER, FS
    [J]. ANALYTICAL BIOCHEMISTRY, 1989, 178 (02) : 408 - 413
  • [4] NEURONAL BIOSENSORS
    BUCH, RM
    RECHNITZ, GA
    [J]. ANALYTICAL CHEMISTRY, 1989, 61 (08) : A533 - +
  • [5] ESFAHANI RN, 1991, THESIS U ILLINOIS CH, P1
  • [6] HOYER LC, 1982, ANTIBODY TOOL APPLIC, P233
  • [7] KASAPBASIOGLU B, 1993, SENSOR ACTUAT B-CHEM, V13, P749
  • [8] KASAPBASIOGLU B, 1992, THESIS U ILLINOIS CH, P1
  • [9] NEWMAN AL, 1986, 2ND P INT M CHEM SEN, V6, P596
  • [10] SANG O, UNPUB