DESIGN OF A GLASS IMPACTOR FOR AN ANNULAR DENUDER FILTER PACK SYSTEM

被引:29
作者
KOUTRAKIS, P
WOLFSON, JM
BRAUER, M
SPENGLER, JD
机构
[1] Harvard School of Public Health, Boston, MA, 02115
关键词
Atmospheric Gases - Denuder Surfaces - Glass Impactors;
D O I
10.1080/02786829008959374
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
A glass impactor for an annular denuder/filter pack system was developedto further the application of denuder technology in sampling atmospheric gases and particles. The glass impactor consists of an entrance section containing the inlet tubethe acceleration jetand the impaction platewhich is mounted at the entrance to the annular denuder. The impaction plate is a removable porous glass disk which can be impregnated with mineral oil to minimize bounce-off of the collected particles during sampling. Calibration tests showed that the impactor has a 50% aerodynamic particle cutoff size of 2.1 μmat a flow of 10 L min−1. Particle loss experiments were conducted. Total losses on surfaces inside the impactorannular denuderand filter packdetermined for particle sizes ranging between 1.50 and 2.77 μmwere lower than 3%. Co-located air sampling was conducted using the glass impactor and the Harvard impactor. Mass concentrations determined using the Harvard impactor were about 10% higher than for the glass impactor because the glass impactor has a slightly lower aerodynamic particle cutoff pointwhile sulfate concentrations obtained from the two systems were in excellent agreement. © 1990 Elsevier Science Publishing Co., Inc.
引用
收藏
页码:607 / 612
页数:6
相关论文
共 9 条