Calibration of electret-based integral radon monitors using NIST polyethylene-encapsulated Ra-226/Rn-222 emanation (PERE) standards

被引:12
作者
Colle, R [1 ]
Kotrappa, P [1 ]
Hutchinson, JMR [1 ]
机构
[1] RAD ELEC INC, FREDERICK, MD 21701 USA
关键词
calibration; electret; emanation; environment; E-PERM(R); ionization chamber; measurement; radium-226; radon-222; standards;
D O I
10.6028/jres.100.047
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The recently developed Rn-222 emanation standards that are based on polyethylene-encapsulated Ra-226 solutions were employed for a first field-measurement application test to demonstrate their efficacy in calibrating passive integral radon monitors. The performance of the capsules was evaluated with respect to the calibration needs of electret ionization chambers (E-PERM(R), Rad Elec Inc.). The encapsulated standards emanate well-characterized and known quantities of Rn-222, and were used in two different-sized, relatively-small, accumulation vessels (about 3.6 L and 10 L) which also contained the deployed electret monitors under test. Calculated integral Rn-222 activities from the capsules over various accumulation times were compared to the averaged electret responses. Evaluations were made with four encapsulated standards ranging in Ra-226 activity from approximately 15 Bq to 540 Bq (with Rn-222 emanation fractions of 0.888); over accumulation times from 1 d to 33 d; and with four different types of E-PERM detectors that were independently calibrated. The ratio of the electret chamber response E(Rn) to the integral Rn-222 activity I-Rn was constant (within statistical variations) over the variables of the specific capsule used, the accumulation volume, accumulation time, and detector type. The results clearly demonstrated the practicality and suitability of the encapsulated standards for providing a simple and readily-available calibration for those measurement applications. However, the mean ratio E(Rn)/I-Rn was approximately 0.91, suggesting a possible systematic bias in the extant E-PERM calibrations. This 9 % systematic difference was verified by an independent test of the E-PERM calibration based on measurements with the NIST radon-in-water standard generator.
引用
收藏
页码:629 / 639
页数:11
相关论文
共 19 条
[1]  
[Anonymous], 1993, GUIDE EXPRESSION UNC
[2]   THE NIST PRIMARY RN-222 MEASUREMENT SYSTEM [J].
COLLE, R ;
HUTCHINSON, JMR ;
UNTERWEGER, MP .
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 1990, 95 (02) :155-165
[3]  
COLLE R, UNPUB PROGR HIST REV
[4]  
COLLE R, 1995, IN PRESS POLYETHYLEN
[5]  
COLLE R, 1995, IN PRESS UPDATE EFFI
[6]  
COLLE R, 1995, IN PRESS NIST POLYET
[7]  
COLLINS R, 1988, GLOBAL TELEVISION, P79
[8]   THE NBS RADON-IN-WATER STANDARD GENERATOR [J].
HUTCHINSON, JMR ;
MULLEN, PA ;
COLLE, R .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 247 (02) :385-389
[9]  
HUTCHINSON JMR, 1992, APPL RADIAT ISOTOPES, V43, P175, DOI 10.1016/0883-2889(92)90090-2
[10]   DEVELOPMENT OF A REGENERATIVE RADON-IN-WATER RADIOACTIVITY STANDARD [J].
HUTCHINSON, JMR ;
MULLEN, PA ;
COLLE, R .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 223 (2-3) :451-457