HIGH-FREQUENCY CURRENT EFFECTS DURING TRANS-URETHRAL RESECTION

被引:22
作者
FLACHENECKER, G
FASTENMEIER, K
机构
关键词
D O I
10.1016/S0022-5347(17)56395-8
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Transurethral resection has been simulated electrically to detect the current pathways and locations of heat production inside the patient. Different electrical behavior was noted with instruments using electrically insulated or metal sheats. Conductive lubricating jelly should be used with the metal sheath and non-conducting jelly should be used with the insulated sheath.
引用
收藏
页码:336 / 341
页数:6
相关论文
共 4 条
[1]  
FLACHENECKER G, 1977, UROLOGE A, V16, P168
[2]  
FLACHENECKER G, 1976, UROLOGE A, V15, P167
[3]   BLADDER EXPLOSIONS DURING TRANSURETHRAL SURGERY [J].
NING, TC ;
ATKINS, DM ;
MURPHY, RC .
JOURNAL OF UROLOGY, 1975, 114 (04) :536-539
[4]  