AN IMPROVEMENT ON THE METHOD OF DETERMINING THE SOLUTE PERMEABILITY OF HOLLOW-FIBER DIALYSIS MEMBRANES PHOTOMETRICALLY USING OPTICAL FIBERS AND COMPARISON OF THE METHOD WITH ORDINARY TECHNIQUES

被引:19
作者
KANAMORI, T [1 ]
SAKAI, K [1 ]
AWAKA, T [1 ]
FUKUDA, M [1 ]
机构
[1] WASEDA UNIV, DEPT CHEM ENGN, 3-4-1 OKUBO, SHINJUKU KU, TOKYO 169, JAPAN
关键词
SOLUTE PERMEABILITY; HOLLOW-FIBER DIALYSIS MEMBRANE; OPTICAL FIBER; VITAMIN-B12; CYTOCHROME-C;
D O I
10.1016/0376-7388(94)87003-9
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
Absorbancy of a solution in the narrow lumen of a tubular membrane under dialysis is continuously measurable with a newly-developed apparatus using quartz optical fibers. The solute permeability of the membrane was determined by calculating time-dependent changes in the absorbancy measured with the apparatus by the mathematical solution derived for unsteady-state concentration profiles in an infinitely long composite cylinder. This method was independent of convective mass transport and osmotic flow through membranes, leading to superiority to ordinary techniques with respect to accuracy.
引用
收藏
页码:159 / 165
页数:7
相关论文
共 9 条
[1]  
Colburn AP, 1933, T AM INST CHEM ENG, V29, P174
[2]  
HOSOYA N, 1990, T AM SOC ART INT ORG, V36, P311
[3]  
KAUFFMANN TG, 1971, AICHE J, V14, P110
[4]   TRANSPORT AND MECHANICAL-PROPERTIES OF HEMODIALYSIS HOLLOW FIBERS [J].
KLEIN, E ;
HOLLAND, F ;
LEBEOUF, A ;
DONNAUD, A ;
SMITH, JK .
JOURNAL OF MEMBRANE SCIENCE, 1976, 1 (04) :371-396
[5]  
MICHAELS AS, 1966, T AM SOC ART INT ORG, V12, P387
[6]  
OHMURA T, 1989, T AM SOC ART INT ORG, V35, P601
[7]   UNSTEADY STATE METHOD FOR MEASURING PERMEABILITY OF SMALL TUBULAR MEMBRANES [J].
STEVENSON, JF ;
VONDEAK, MA ;
WEINBERG, M ;
SCHUETTE, RW .
AICHE JOURNAL, 1975, 21 (06) :1192-1199
[8]   UNSTEADY MASS-TRANSFER IN A LONG COMPOSITE CYLINDER WITH INTERFACIAL RESISTANCES [J].
STEVENSON, JF .
AICHE JOURNAL, 1974, 20 (03) :461-466
[9]  
Stiller S, 1985, Contrib Nephrol, V46, P23