A MULTIPLEXED SINGLE-PHOTON INSTRUMENT FOR ROUTINE MEASUREMENT OF TIME-RESOLVED FLUORESCENCE ANISOTROPY

被引:24
作者
BIRCH, DJS [1 ]
HOLMES, AS [1 ]
GILCHRIST, JR [1 ]
IMHOF, RE [1 ]
ALALAWI, SM [1 ]
NADOLSKI, B [1 ]
机构
[1] EDINBURGH INSTRUMENTS LTD,EDINBURGH EH14 4AP,SCOTLAND
来源
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS | 1987年 / 20卷 / 04期
关键词
OPTICAL INSTRUMENTS - Design;
D O I
10.1088/0022-3735/20/4/031
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We report a new type of instrument, the Edinburgh Instruments Model 299T differential fluorometer, for the measurement of the decay of fluorescence anisotropy. The instrument incorporates three multiplexed detection channels for simultaneous measurement of two polarized fluorescence decays and the excitation pulse, using time-correlated single-photon counting. This arrangement corrects automatically for temporal and intensity fluctuations of the spark source. A study of the probe 1,6-diphenyl-1,3,5-hexatriene (DPH) in a white oil reveals that the instrument can measure rotational correlation times with subnanosecond time resolution in a third of the time of conventional methods.