THE EFFECT OF CORNEAL THICKNESS ON NONCONTACT TONOMETRY

被引:24
作者
GRAF, M
机构
关键词
D O I
10.1055/s-2008-1046068
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
In this study the Reichert non-contact tonometer (NCT II) was compared with the Goldmann applanation tonometer (GAT) in 25 subjects (50 eyes; pressure range 9-17 mmHg). The relationship NCT/GAT fulfilled the conditions required for legal verification. Marked differences between the readings of both instruments can be caused by anomalous central corneal thickness, because non-contact tonometry is influenced more than conventional applanation tonometry by the corneal thickness. Starting from a central thickness of 0.51 mm, the NCT is assumed to underestimate the actual intraocular pressure in eyes with thinner and to overestimate it in eyes with thicker cornea by 1 mmHg per 0.01 mm difference of corneal thickness. This value changes with individual corneal tissue qualities and is not valid especially in presence of corneal edema.
引用
收藏
页码:183 / 186
页数:4
相关论文
共 17 条
[1]  
Clemens S, 1988, Fortschr Ophthalmol, V85, P154
[2]  
DRAEGER J, 1975, KLIN MONATSBL AUGENH, V167, P27
[3]  
DRAEGER J, 1976, GLAUKOM S WURZBURG, P183
[4]  
EHLERS N, 1975, ACTA OPHTHALMOL, V53, P34
[5]  
FRIEDENWALD J S, 1957, Trans Am Acad Ophthalmol Otolaryngol, V61, P108
[6]  
GOLDMANN H, 1957, Ophthalmologica, V134, P221
[7]   PREOPERATIVE AND POSTOPERATIVE NONCONTACT TONOMETRY [J].
GRAF, M ;
HESSEMER, V ;
JACOBI, KW .
KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1990, 197 (02) :182-186
[8]  
GROLMAN B, 1972, AMER J OPT ARCH AM A, V49, P646
[9]   NOTES ON CLINICAL-CRITERIA FOR THE EVALUATION OF TONOMETERS [J].
JESSEN, K ;
HOFFMANN, F .
KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1982, 180 (01) :96-99
[10]   CLINICAL AND STATISTICAL ASPECTS ON STANDARDIZATION OF TONOMETERS [J].
JESSEN, K ;
LUEBBIG, H ;
WEIGELIN, E .
ALBRECHT VON GRAEFES ARCHIV FUR KLINISCHE UND EXPERIMENTELLE OPHTHALMOLOGIE, 1979, 209 (04) :269-278