AN EMISSION-BASED FEEDBACK-SYSTEM FOR STABILIZING AN INDUCTIVELY COUPLED PLASMA

被引:9
作者
MARKS, MA [1 ]
HIEFTJE, GM [1 ]
机构
[1] INDIANA UNIV,DEPT CHEM,BLOOMINGTON,IN 47405
关键词
ARGON; -; SPECTROSCOPY; EMISSION; Applications;
D O I
10.1366/0003702884428112
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An emission-controlled feedback system for stabilizing an ICP (Inductively Coupled Plasma) has been investigated. Unlike previous stabilization methods which monitor only the power being sent to the plasma, the present one uses an emission signal from either argon or an internal reference element as a stability indicator. Using a previously constructed power-modulation circuit for rf power control, the feedback network adjusts the plasma power to compensate for changes in the emission level of the reference element. It is shown to be particularly useful for reducing or eliminating the drift that occurs when the plasma is warming up. Results with several combinations of argon, barium, calcium, iron, manganese, sodium, and strontium lines are presented.
引用
收藏
页码:277 / 280
页数:4
相关论文
共 10 条
[1]  
HARRISON GR, 1939, MIT WAVELENGTH TABLE
[2]  
Meggers W.F., 1975, NBS MONOGRAPH, V145
[3]  
Moore C. E, 1971, NSRDS NBS, V35
[4]  
RAYSON GD, 1985, PITTSBURGH C ANAL CH
[5]  
RAYSON GD, 1985, A C S NATIONAL M CHI
[6]  
RAYSON GD, 1985, ANN ROCKY MOUNTAIN C
[7]   CORRECTION OF MATRIX EFFECTS IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY BY INTERACTIVE POWER ADJUSTMENT [J].
THOMPSON, M ;
RAMSEY, MH ;
COLES, BJ ;
DU, CM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (02) :185-188
[8]  
WEISE WL, 1980, NSRDS NBS, V68
[9]   MODIFICATION OF AN INDUCTIVELY COUPLED PLASMA RADIO-FREQUENCY SUPPLY FOR AMPLITUDE-MODULATION WITH COMPLEX WAVE FORMS [J].
WITHNELL, R ;
RAYSON, GD ;
PARISI, AF ;
HIEFTJE, GM .
ANALYTICAL CHEMISTRY, 1985, 57 (12) :2414-2417
[10]  
1975, CRC HDB CHEM PHYSICS