ABSOLUTE CALIBRATION OF A FLAME PHOTOMETRIC DETECTOR TO VOLATILE SULFUR COMPOUNDS AT SUB-PART-PER-MILLION LEVELS

被引:77
作者
STEVENS, RK
OKEEFFE, AE
ORTMAN, GC
机构
[1] National Air Pollution Control Administration, Cincinnati
关键词
D O I
10.1021/es60030a005
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Permeation tubes were used to obtain an absolute calibration of a flame photometric detector (FPD) to sulfur dioxide, hydrogen sulfide, methyl mercaptan, and carbondisulfide. Response data was obtained from 0.001 to 10 p.p.m. for each compound. Preliminary experiments with the FPD as anambient air sulfur dioxide monitor indicated that the detector has potential utility as a continuous monitoring device. © 1969, American Chemical Society. All rights reserved.