ALTERATION OF PULMONARY BLOOD-FLOW BY PULMONARY-ARTERY OCCLUDED PRESSURE MEASUREMENT

被引:7
作者
BERRY, AJ [1 ]
GEER, RT [1 ]
MARSHALL, BE [1 ]
机构
[1] UNIV PENN,SCH MED,DEPT ANESTHESIOL,PHILADELPHIA,PA 19104
关键词
D O I
10.1097/00000542-197908000-00017
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
It is demonstrated that pulmonary arterial blood flow can be significantly reduced in certain patients by Swan-Ganz catheter balloon inflation, with resultant decreases in SAP and cardiac output and increase in CVP. As pulmonary blood flow is reduced, left atrial pressure and PAOP as measured are decreased from preocclusion values, and may lead to inappropriate fluid or vasopressor therapy and to errors in values calculated from PAOP.
引用
收藏
页码:164 / 166
页数:3
相关论文
共 8 条
[1]  
CHUN GMH, 1971, NEW ENGL J MED, V284, P1041
[2]  
COURNAND A, 1950, J THORAC SURG, V19, P80
[3]   PULMONARY COMPLICATIONS OF FLOW-DIRECTED BALLOON-TIPPED CATHETER [J].
FOOTE, GA ;
SCHABEL, SI ;
HODGES, M .
NEW ENGLAND JOURNAL OF MEDICINE, 1974, 290 (17) :927-931
[4]   EVALUATION OF PULMONARY ARTERIAL END-DIASTOLIC PRESSURE AS AN INDIRECT ESTIMATE OF LEFT ATRIAL MEAN PRESSURE [J].
JENKINS, BS ;
BRADLEY, RD ;
BRANTHWAITE, MA .
CIRCULATION, 1970, 42 (01) :75-+
[5]   HEMOPTYSIS WITH FLOW-DIRECTED CARDIAC CATHETERIZATION [J].
LAPIN, ES ;
MURRAY, JA .
JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1972, 220 (09) :1246-&
[6]   INDIRECT MEASUREMENT OF LEFT-ATRIAL PRESSURE IN SURGICAL PATIENTS - PULMONARY-CAPILLARY WEDGE AND PULMONARY-ARTERY DIASTOLIC PRESSURES COMPARED WITH LEFT-ATRIAL PRESSURE [J].
LAPPAS, D ;
LELL, WA ;
GABEL, JC ;
CIVETTA, JM ;
LOWENSTEIN, E .
ANESTHESIOLOGY, 1973, 38 (04) :394-397
[7]   PITFALLS OF SWAN-GANZ CATHETERIZATION [J].
SHIN, B ;
AYELLA, RJ ;
MCASLAN, TC .
CRITICAL CARE MEDICINE, 1977, 5 (03) :125-127
[8]   COMPARISON OF PULMONARY WEDGE AND LEFT ATRIAL PRESSURE IN MAN [J].
WALSTON, A ;
KENDALL, ME .
AMERICAN HEART JOURNAL, 1973, 86 (02) :159-164