EFFECTS OF MOISTURE ON THE COLD VAPOR DETERMINATION OF MERCURY AND ITS REMOVAL BY USE OF MEMBRANE DRYER TUBES

被引:32
作者
CORNS, WT [1 ]
EBDON, L [1 ]
HILL, SJ [1 ]
STOCKWELL, PB [1 ]
机构
[1] PS ANALYT LTD,SEVENOAKS TN15 6QY,KENT,ENGLAND
关键词
VAPOR GENERATION; ATOMIC FLUORESCENCE SPECTROMETRY; SEMI-PERMEABLE MEMBRANE DRYER TUBE; MERCURY DETERMINATION; MOISTURE REMOVAL;
D O I
10.1039/an9921700717
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Continuous-flow vapour generation is now a well-established sample-introduction technique for mercury and the hydride-forming elements. One of the problems associated with this technique is moisture carryover, which originates during the gas-liquid separation process. Excessive moisture carryover causes gradual loss in sensitivity and baseline drift for atomic absorption and atomic fluorescence spectrometry and inductively coupled plasma detection systems. Chemical desiccants and physical moisture traps can be used to reduce moisture for short periods of time, although these may give rise to contamination and analyte losses. A more effective way to remove moisture carryover has been found by using a semi-permeable Nafion membrane dryer tube, which continuously desolvates the wet gaseous stream. This device improved the long-term stability and enhanced sensitivity. A relative standard deviation of 2% was achieved for 90 runs of a 1-mu-g l-1 standard, obtained over a period of 3 h.
引用
收藏
页码:717 / 719
页数:3
相关论文
共 16 条