共 3 条
DTT quenches the passive reference signal in real-time PCR
被引:10
作者:
Boiso, Lina
[1
]
Sanga, Malin
[1
]
Hedman, Johannes
[1
,2
]
机构:
[1] Swedish Natl Forens Ctr, SE-58194 Linkoping, Sweden
[2] Lund Univ, Dept Chem, Appl Microbiol, Lund, Sweden
关键词:
Dithiothreitol;
Fluorescence quenching;
Passive reference dye;
PCR inhibition;
qPCR;
D O I:
10.1016/j.fsigss.2015.09.003
中图分类号:
Q3 [遗传学];
学科分类号:
071007 ;
090102 ;
摘要:
Forensic DNA analysis is partly limited by PCR-inhibitory compounds present in the DNA extracts. Generally, these inhibitors disturb amplification, i.e., the production of amplicons. We have found that dithiothreitol (DTT) from the DNA extraction process can cause another type of real-time PCR disturbance, i.e., inhibition of signal detection through fluorescence quenching. DNA extracts containing DTT substantially quenched the passive reference signal in the Quantifiler HP DNA Quantification kit. This quenching resulted in overestimation of DNA concentrations, as target DNA signals are normalized to the passive reference signal. (C) 2015 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:E5 / E6
页数:2
相关论文