The potentiometric response to anions of glassy carbon electrodes coated with thin unmetallated poly(tetrakis(p-aminophenyl)porphyrin) (poly(H2(p-TAPP))) films is examined. The polymer films are introduced onto the glassy carbon surface by constant-potential anodic electrolysis of the porphyrin monomer from a solution of dichloromethane containing a quaternary ammonium perchlorate salt. The thickness of the film is controlled by the electrodeposition time. Oxidized poly(porphyrin) modified electrodes exhibit near Nernstian response to iodide in the concentration range of 5 x 10(-5) - 1 x 10(-1) M with significant selectivity over a wide range of other anions tested. The mechanism of the electrode's iodide ion response is considered in light of its response to other redox-active anions, and a recent literature report describing the selective iodide response of electrodes coated with electropolymerized poly(3-methylthiophene) films. Spectroelectrochemical studies with poly(porphyrin)-modified tin oxide electrodes are used to confirm the redox-type response mechanism of the sensors toward iodide. The analytical limitations of using such polymer film-modified electrodes for potentiometric ion sensing are discussed.
引用
收藏
页码:841 / 849
页数:9
相关论文
共 30 条
[1]   ANION SELECTIVITY OF METALLOPORPHYRINS IN MEMBRANES [J].
AMMANN, D ;
HUSER, M ;
KRAUTLER, B ;
RUSTERHOLZ, B ;
SCHULTHESS, P ;
LINDEMANN, B ;
HALDER, E ;
SIMON, W .
HELVETICA CHIMICA ACTA, 1986, 69 (04) :849-854
[2]   ELECTROCHEMICAL POLYMERIZATION OF AMINO-SUBSTITUTED, PYRROLE-SUBSTITUTED, AND HYDROXY-SUBSTITUTED TETRAPHENYLPORPHYRINS [J].
BETTELHEIM, A ;
WHITE, BA ;
RAYBUCK, SA ;
MURRAY, RW .
INORGANIC CHEMISTRY, 1987, 26 (07) :1009-1017
[3]   MN(III) PORPHYRIN-BASED THIOCYANATE-SELECTIVE MEMBRANE ELECTRODES - CHARACTERIZATION AND APPLICATION IN FLOW-INJECTION DETERMINATION OF THIOCYANATE IN SALIVA [J].
BROWN, DV ;
CHANIOTAKIS, NA ;
LEE, IH ;
MA, SC ;
PARK, SB ;
MEYERHOFF, ME ;
NICK, RJ ;
GROVES, JT .
ELECTROANALYSIS, 1989, 1 (06) :477-484
[4]  
CATTRALL RW, 1984, ION SEL ELECTRODE R, V6, P125
[5]   INFLUENCE OF PORPHYRIN STRUCTURE ON ANION SELECTIVITIES OF MANGANESE(III) PORPHYRIN BASED MEMBRANE ELECTRODES [J].
CHANIOTAKIS, NA ;
CHASSER, AM ;
MEYERHOFF, ME ;
GROVES, JT .
ANALYTICAL CHEMISTRY, 1988, 60 (02) :185-188
[6]   SALICYLATE-SELECTIVE MEMBRANE-ELECTRODE BASED ON TIN(IV) TETRAPHENYLPORPHYRIN [J].
CHANIOTAKIS, NA ;
PARK, SB ;
MEYERHOFF, ME .
ANALYTICAL CHEMISTRY, 1989, 61 (06) :566-570
[7]   ANION-SELECTIVE ELECTRODES BASED ON ELECTROPOLYMERIZED PORPHYRIN FILMS [J].
DAUNERT, S ;
WALLACE, S ;
FLORIDO, A ;
BACHAS, LG .
ANALYTICAL CHEMISTRY, 1991, 63 (17) :1676-1679
[8]  
DONG S, 1988, J CHEM SOC CHEM COMM, V105, P993
[9]   THE APPLICATION OF CHEMICALLY MODIFIED ELECTRODES IN ANALYTICAL-CHEMISTRY [J].
DONG, SJ ;
WANG, YD .
ELECTROANALYSIS, 1989, 1 (02) :99-106
[10]   CHLORIDE CHEMICAL SENSOR BASED ON AN ORGANIC CONDUCTING POLYPYRROLE POLYMER [J].
DONG, SJ ;
SUN, ZS ;
LU, ZL .
ANALYST, 1988, 113 (10) :1525-1528