RESISTANCE TO CONSTANT AIR-FLOW IMPOSED BY THE STANDARD LARYNGEAL MASK, THE REINFORCED LARYNGEAL MASK AND RAE TRACHEAL TUBES

被引:11
作者
ALHASANI, A
机构
[1] Division of Anaesthesia, Leicester Royal Infirmary
关键词
EQUIPMENT; LARYNGEAL MASK AIRWAY; REINFORCED LARYNGEAL MASK; AIRWAY; RESISTANCE;
D O I
10.1093/bja/71.4.594
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The reinforced laryngeal mask airway (RLMA) has been introduced recently for head and neck surgery. Its resistance to constant air flow has been measured and compared with that of the standard laryngeal mask airway (LMA). The RLMA resistance has also been compared with that imposed by the standard oral Ring-Adair-Elwyn (RAE) tube of a corresponding size for a given patient. The resistance to gas flow of the new RLMA was approximately three to five times that of the LMA at two flow rates (0.5 and 1.0 litre s-1). The resistance of the RLMA 4 was intermediate between that of RAE tubes sizes 8 and 9 mm, and that of RLMA 2 intermediate between that of RAE tubes sizes 5 and 6 mm.
引用
收藏
页码:594 / 596
页数:3
相关论文
共 4 条
[1]   RESISTANCE AND ADDITIONAL INSPIRATORY WORK IMPOSED BY THE LARYNGEAL MASK AIRWAY - A COMPARISON WITH TRACHEAL TUBES [J].
BHATT, SB ;
KENDALL, AP ;
LIN, ES ;
OH, TE .
ANAESTHESIA, 1992, 47 (04) :343-347
[2]  
DEMERS RR, 1977, JAMA-J AM MED ASSOC, V237, P1362, DOI 10.1001/jama.1977.03270400066026
[3]   AIRWAY PROTECTION BY THE LARYNGEAL MASK - A BARRIER TO DYE PLACED IN THE PHARYNX [J].
JOHN, RE ;
HILL, S ;
HUGHES, TJ .
ANAESTHESIA, 1991, 46 (05) :366-367
[4]  
1992, MANUAL FLEXIBLE REIN