RADIOFREQUENCY MAP OF AN NMR COIL BY IMAGING

被引:134
作者
AKOKA, S
FRANCONI, F
SEGUIN, F
LEPAPE, A
机构
[1] Laboratoire de Biophysique Cellulaire, RMN, Faculté de Médecine de Tours, 37032 Tours Cedex, 2bis, boulevard Tonnellé
关键词
B-1; DISTRIBUTION; RADIOFREQUENCY MAP;
D O I
10.1016/0730-725X(93)90078-R
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We propose a new imaging method to obtain a map of the radiofrequency (RF) field amplitude over a sample. The sequence contains three RF pulses (alpha, 2alpha, and alpha) and produces two images by a classical spin echo and a stimulated echo. A third image is computed and gives the distribution of the flip angle a, and so the RF amplitude, over the sample. The accuracy of the flip angle determination is verified on an homogeneous sample and results show a good correlation between experimental and theoretical flip angles in the range of 50-degrees to 130-degrees. Experiments with a surface coil and a resonator show the method is available in an inhomogeneous RF field. Images obtained on the calf of a volunteer confirms the independence of the computed RF distribution from proton density, T1, or T2 contrast.
引用
收藏
页码:437 / 441
页数:5
相关论文
共 6 条
  • [1] Hahn, Spin echoes, Phys. Rev., 80, pp. 580-594, (1950)
  • [2] Perman, Bernstein, Sandstrom, A method for correctly setting the rf flip angle, Magn. Reson. Med., 9, pp. 16-24, (1989)
  • [3] Carlson, Kramer, Rapid radiofrequency calibration in MRI, Magn. Reson. Med., 15, pp. 438-445, (1990)
  • [4] Graumann, Oppelt, Stetler, Multiple-spin-echo imaging with a 2D Fourier method, Magn. Reson. Med., 3, pp. 707-721, (1986)
  • [5] Roschmann, Radiofrequency penetration and absorption in the human body: limitations to high-field whole-body nuclear magnetic resonance imaging, Med. Phys., 14, pp. 922-931, (1987)
  • [6] Young, Bryant, Payne, Variations in slice shape and absorption as artefacts in the determination of tissues parameters in NMR imaging, Magn. Reson. Med., 2, pp. 355-389, (1985)