FLUOROMETRIC-DETERMINATION OF POTASSIUM-ION USING HEXADECYL-ACRIDINE ORANGE IMMOBILIZED ON A POLY(VINYL CHLORIDE) MEMBRANE ATTACHED TO A FLOW-THROUGH CELL

被引:7
作者
KAWABATA, Y
IMASAMA, T
ISHIBASHI, N
机构
[1] Faculty of Engineering, Kyushu University, Fukuoka, 812, Hakozaki
关键词
FLOW SYSTEM; FLUOROMETRY; BLOOD; MEMBRANE ELECTRODES; POTASSIUM;
D O I
10.1016/0003-2670(91)85092-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A cationic fluorescent probe of hexadecyl-acridine orange is immobilized on a poly(vinyl chloride) membrane containing valinomycin. The membrane is attached to the inner wall of the flow-through cell in a fluorescence spectrometer, and potassium ion is detected repetitively by a flow injection system without consuming the indicator reagent. The response is proportional to the logarithmic concentration of the potassium ion in the range 0.5 - 50 mM. Fifteen samples are determined in 1 h, and the relative standard deviation of the peak height is 2.4% for a sample containing 0.1-mu-mol (1 mM) of potassium ion. The extended Nicolsky equation is applied to the observed peak height, the selectivity coefficient for sodium ion being 4 x 10(-3). The concentration of potassium ion in human blood plasma is also determined by this system.
引用
收藏
页码:97 / 101
页数:5
相关论文
共 16 条
[1]  
BERND AA, 1989, ANAL CHIM ACTA, V225, P293
[2]   OPTICAL SENSOR FOR CALCIUM - PERFORMANCE, STRUCTURE, AND REACTIVITY OF CALCICHROME IMMOBILIZED AT AN ANIONIC POLYMER FILM [J].
CHAU, LK ;
PORTER, MD .
ANALYTICAL CHEMISTRY, 1990, 62 (18) :1964-1971
[3]   CHEMICAL SENSORS [J].
JANATA, J ;
BEZEGH, A .
ANALYTICAL CHEMISTRY, 1988, 60 (12) :R62-R74
[4]   FIBEROPTIC POTASSIUM-ION SENSOR USING ALKYL-ACRIDINE ORANGE IN PLASTICIZED POLY(VINYL CHLORIDE) MEMBRANE [J].
KAWABATA, Y ;
TAHARA, R ;
KAMICHIKA, T ;
IMASAKA, T ;
ISHIBASHI, N .
ANALYTICAL CHEMISTRY, 1990, 62 (14) :1528-1531
[5]   ION-SELECTIVE OPTRODE USING HEXADECYL-ACRIDINE ORANGE ATTACHED ON POLY(VINYL CHLORIDE) MEMBRANE [J].
KAWABATA, Y ;
KAMICHIKA, T ;
IMASAKA, T ;
ISHIBASHI, N .
ANALYTICAL CHEMISTRY, 1990, 62 (18) :2054-2055
[6]  
MALINSKI T, 1991, APPL FLUORESCENCE TE, V3, P1
[7]  
MORF WE, 1978, ION SELECTIVE ELECTR, V1, P270
[8]  
MORF WE, 1978, ION SELECTIVE ELECTR, V1, P216
[9]  
RUZICKA J, 1981, FLOW INJECTION ANAL, P21
[10]   A CALCIUM-SELECTIVE OPTRODE BASED ON FLUORIMETRIC MEASUREMENT OF MEMBRANE-POTENTIAL [J].
SCHAFFAR, BPH ;
WOLFBEIS, OS .
ANALYTICA CHIMICA ACTA, 1989, 217 (01) :1-9