DETERMINATION OF THE LOCAL TEMPERATURE AT A SAMPLE DURING RAMAN EXPERIMENTS USING STOKES AND ANTI-STOKES RAMAN BANDS

被引:83
作者
KIP, BJ [1 ]
MEIER, RJ [1 ]
机构
[1] DSM RES & PATENTS,POB 18,6160 MD GELEEN,NETHERLANDS
关键词
Anti-Stokes Raman Bands - Stokes Raman Bands - Stokes/Anti-Stokes Intensity Ratio;
D O I
10.1366/0003702904087325
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Sample heating due to high incident laser power in Raman experiments on (industrial) polymers and catalysts can easily lead to unwanted changes in morphology or desorption of surface species. Therefore temperature of the sample at the laser spot was studied by determining Stokes and anti-Stokes Raman intensities. Temperature can normally be determined within an accuracy of about 5 K, although for samples having nothing but Raman bands above 1000 cm-1, the anti-Stokes lines are very weak and accuracy of temperature determination decreased drastically. For most of the samples studied in the macro-Raman setup, powers up to 250 mW could be used without measurable heating. In the micro-Raman setup heating started earlier due to higher focusing.
引用
收藏
页码:707 / 711
页数:5
相关论文
共 5 条
[1]   RESONANCE RAMAN-SPECTRUM OF DEGRADED POLYVINYL-CHLORIDE) .2. GAMMA-IRRADIATED SAMPLES [J].
GERRARD, DL ;
MADDAMS, WF .
MACROMOLECULES, 1977, 10 (06) :1221-1224
[2]   RESONANCE RAMAN-SPECTRUM OF THERMALLY DEGRADED POLYVINYL-CHLORIDE) [J].
GERRARD, DL ;
MADDAMS, WF .
MACROMOLECULES, 1975, 8 (01) :54-58
[3]  
Hayes W., 1978, SCATTERING LIGHT CRY, P26
[4]  
KONINGSTEIN JA, 1972, INTRO THEORY RAMAN E, P106
[5]  
Long D. A., 1976, RAMAN SPECTROSCOPY, P84