SURFACE DAMAGE TO COBALT-CHROME FEMORAL-HEAD PROSTHESES

被引:83
作者
JASTY, M [1 ]
BRAGDON, CR [1 ]
LEE, K [1 ]
HANSON, A [1 ]
HARRIS, WH [1 ]
机构
[1] HARVARD UNIV,SCH MED,BOSTON,MA
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1994年 / 76B卷 / 01期
关键词
D O I
10.1302/0301-620X.76B1.8300686
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Fifty-four cobalt-chrome alloy femoral heads were retrieved at revision surgery or at post-mortem; 23 came from uncemented, eight from hybrid and 23 from cemented hip arthroplasties. The uncemented and hybrid implants had porous coating, metal backing and modular femoral heads; with one exception none of the cemented implants had any of these. Twenty-five of the 31 heads from uncemented and hybrid arthroplasties, and 11 of the 23 heads from cemented arthroplasties showed surface damage involving more than 25% of the surface. Scanning electron microscopy revealed multidirectional fine scratches 1 mu m to 10 mu m in depth and width which appeared to have been made by fine, hard particles. There was a higher rate of such damage in the uncemented and hybrid arthroplasties than in the cemented implants, suggesting that the abrasive particles were mainly released from the metal, rather than from the cement or polyethylene components of the implants.
引用
收藏
页码:73 / 77
页数:5
相关论文
共 21 条
[1]   METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS [J].
AGINS, HJ ;
ALCOCK, NW ;
BANSAL, M ;
SALVATI, EA ;
WILSON, PD ;
PELLICCI, PM ;
BULLOUGH, PG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :347-356
[2]   LOCALIZED ENDOSTEAL BONE LYSIS IN RELATION TO THE FEMORAL COMPONENTS OF CEMENTED TOTAL HIP ARTHROPLASTIES [J].
ANTHONY, PP ;
GIE, GA ;
HOWIE, CR ;
LING, RSM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (06) :971-979
[3]  
Caravia L, 1990, Proc Inst Mech Eng H, V204, P65, DOI 10.1243/PIME_PROC_1990_204_230_02
[4]   FORMATION OF A SYNOVIAL-LIKE MEMBRANE AT THE BONE-CEMENT INTERFACE - ITS ROLE IN BONE-RESORPTION AND IMPLANT LOOSENING AFTER TOTAL HIP-REPLACEMENT [J].
GOLDRING, SR ;
JASTY, M ;
ROELKE, MS ;
ROURKE, CM ;
BRINGHURST, FR ;
HARRIS, WH .
ARTHRITIS AND RHEUMATISM, 1986, 29 (07) :836-842
[5]   EXTENSIVE LOCALIZED BONE-RESORPTION IN FEMUR FOLLOWING TOTAL HIP-REPLACEMENT [J].
HARRIS, WH ;
SCHILLER, AL ;
SCHOLLER, JM ;
FREIBERG, RA ;
SCOTT, R .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1976, 58 (05) :612-618
[6]  
Isaac G H, 1987, Eng Med, V16, P167
[7]  
Isaac G H, 1986, Eng Med, V15, P19
[8]   LOCALIZED OSTEOLYSIS IN STABLE, NONSEPTIC TOTAL HIP-REPLACEMENT [J].
JASTY, MJ ;
FLOYD, WE ;
SCHILLER, AL ;
GOLDRING, SR ;
HARRIS, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (06) :912-919
[9]  
Kurth M, 1988, J Biomater Appl, V3, P33, DOI 10.1177/088532828800300102
[10]   ASEPTIC LOOSENING IN TOTAL HIP-ARTHROPLASTY SECONDARY TO OSTEOLYSIS INDUCED BY WEAR DEBRIS FROM TITANIUM-ALLOY MODULAR FEMORAL HEADS [J].
LOMBARDI, AV ;
MALLORY, TH ;
VAUGHN, BK ;
DROUILLARD, P .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1989, 71A (09) :1337-1342