DIRECT INJECTION NEBULIZATION FOR INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY

被引:154
作者
WIEDERIN, DR
SMITH, FG
HOUK, RS
机构
[1] IOWA STATE UNIV SCI & TECHNOL,US DOE,AMES LAB,AMES,IA 50011
[2] IOWA STATE UNIV SCI & TECHNOL,DEPT CHEM,AMES,IA 50011
关键词
D O I
10.1021/ac00003a007
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An improved direct injection nebulizer (DIN) is used in a continuous-flow mode for inductively coupled plasma mass spectrometry (ICP/MS). Precision in cases where the signal is not limited by counting statistics is typically 0.5% relative standard deviation (RSD), an improvement by a factor of 4 over the precision obtained with conventional nebulizers. A rinse-out time of only 15 s is sufficient to reduce the analyte signal to 0.1% of its steady-state value, even for memory-prone elements such as Hg and I. Relative detection limits are similar to those obtained with conventional pneumatic nebulizers. Since the sample flow rate is only 120-mu-L min-1, the absolute detection limits are improved by up to an order of magnitude. The refractory oxide ion ratios (MO+/M+) are 3 times higher than with other nebulizers. Matrix effects caused by Na are comparable to those observed with other nebulizers.