A CLINICAL-EVALUATION OF 5 ELECTRONIC ROOT-CANAL LENGTH MEASURING-INSTRUMENTS

被引:75
作者
FOUAD, AF
KRELL, KV
MCKENDRY, DJ
KOORBUSCH, GF
OLSON, RA
机构
[1] SUNY BUFFALO,SCH DENT MED,DEPT ENDODONT,BUFFALO,NY 14260
[2] UNIV IOWA,COLL DENT,DEPT ORAL & MAXILLOFACIAL SURG,IOWA CITY,IA 52242
关键词
D O I
10.1016/S0099-2399(06)81889-3
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
A previous in vitro study has shown high accuracy, but no clinically significant differences in a group of five electronic root canal length measuring instruments. The purpose of this study was to evaluate and compare the performance of the same group of instruments under clinical conditions and to correlate their accuracy to radiographic estimates of canal length. Five electronic root canal length measuring instruments were used to measure the working length to the "apex" in 20 single-rooted teeth scheduled for extraction. After extraction, the actual canal length was measured visually to a point just within the apical foramen. This length was compared with instrument length as determined electronically. The accuracy of the instruments in determining canal measurement with ±0.5 mm from the apical foramen varied from 55 to 75%. The differences between the instruments were not statistically significant. On average, all of the instruments except for the Endocater gave canal length measurements that were beyond the apical foramen. The variability of the measurements, which was comparable to that of estimates of canal length from preoperative radiographs, indicated that radiographic verification of the working length is still desirable. © 1990 The American Association of Endodontists.
引用
收藏
页码:446 / 449
页数:4
相关论文
共 20 条
[1]   A MODEL FOR DEMONSTRATING AN ELECTRONIC CANAL LENGTH MEASURING DEVICE [J].
AURELIO, JA ;
NAHMIAS, Y ;
GERSTEIN, H .
JOURNAL OF ENDODONTICS, 1983, 9 (12) :568-569
[2]   EVALUATION OF THE ACCURACY OF THE NEOSONO-D ELECTRONIC APEX LOCATOR [J].
BERMAN, LH ;
FLEISCHMAN, SB .
JOURNAL OF ENDODONTICS, 1984, 10 (04) :164-167
[3]  
Blank L W, 1975, J Endod, V1, P141, DOI 10.1016/S0099-2399(75)80099-9
[4]   CRITICAL EVALUATION OF SOME METHODS OF DETERMINING TOOTH LENGTH [J].
BRAMANTE, CM ;
BERBERT, A .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1974, 37 (03) :463-473
[5]  
Busch L R, 1976, J Endod, V2, P295, DOI 10.1016/S0099-2399(76)80043-X
[6]  
Cash P W, 1975, J Am Dent Assoc, V91, P1135
[7]   INVIVO ROOT-CANAL LENGTH DETERMINATION USING THE FORAMETER [J].
CHUNN, CB ;
ZARDIACKAS, LD ;
MENKE, RA .
JOURNAL OF ENDODONTICS, 1981, 7 (11) :515-520
[8]  
England M C, 1987, Va Dent J, V64, P20
[9]   AN INVITRO COMPARISON OF 5 ROOT-CANAL LENGTH MEASURING-INSTRUMENTS [J].
FOUAD, AF ;
KRELL, KV .
JOURNAL OF ENDODONTICS, 1989, 15 (12) :573-577
[10]  
HASEGAWA K, 1986, Journal of Nihon University School of Dentistry, V28, P117