ULTRASENSITIVE DETERMINATION OF EUROPIUM USING MICROSECOND TIME-RESOLVED SPECTROFLUORIMETRY

被引:10
作者
CHRISTOPOULOS, TK
DIAMANDIS, EP
机构
[1] TORONTO WESTERN HOSP,DEPT CLIN BIOCHEM,399 BATHURST ST,TORONTO M5T 2S8,ONTARIO,CANADA
[2] UNIV TORONTO,DEPT CLIN BIOCHEM,TORONTO M5G 1L5,ONTARIO,CANADA
关键词
LANTHANIDE-ION CHELATES; TIME-RESOLVED SPECTROFLUORIMETRY; EUROPIUM; ULTRASENSITIVE ASSAY; TRACE METAL DETERMINATION;
D O I
10.1039/an9911600627
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
4,7-Bis(chlorosulphophenyl)-1,10-phenanthroline-2,9-dicarboxylic acid (BCPDA) is proposed as a simple, highly sensitive and specific reagent for the ultrasensitive determination of Eu3+ with microsecond time-resolved spectrofluorimetry. The optimum reaction pH was 9.0 in a glycine buffer. Detection limit for EU3+ was 3 x 10(-13) mol l-1. No major interferences from other metals were found when the interferent was present in a 500-fold molar excess. Negative interference was observed at high (> 1 X 10(-3) mol l-1) concentrations of Fe3+ and Ca2+ presumably through the formation of non-fluorescent complexes with BCPDA.
引用
收藏
页码:627 / 630
页数:4
相关论文
共 17 条