COMPARISON OF DEFIBRILLATION PROBABILITY OF SUCCESS CURVES FOR AN ENDOCARDIAL LEAD CONFIGURATION WITH AND WITHOUT AN INACTIVE EPICARDIAL PATCH

被引:10
作者
CALLIHAN, RL
IDRISS, SF
DAHL, RW
WOLF, PD
SMITH, WM
IDEKER, RE
机构
[1] DUKE UNIV, MED CTR, DEPT MED, DURHAM, NC 27710 USA
[2] DUKE UNIV, MED CTR, DEPT PATHOL, DURHAM, NC 27710 USA
[3] DUKE UNIV, SCH ENGN, DEPT BIOMED ENGN, DURHAM, NC 27706 USA
[4] DUKE UNIV, ENGN RES CTR EMERGING CARDIOVASC TECHNOL, DURHAM, NC 27706 USA
关键词
D O I
10.1016/0735-1097(94)00558-8
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. This study sought to assess the effect of passive ''bystander'' epicardial electrodes on defibrillation efficacy. Background. We hypothesized that an inactive epicardial patch placed in an area of low potential gradient from an endocardial electrode shock might affect defibrillation efficacy through its effects on the shock field and the underlying potential gradient. Methods. We studied the effects of an inactive 18-cm(2) titanium mesh patch placed on the anterolateral left ventricular epicardium on the 50% probability of successful defibrillation. A biphasic shock with both phases 6 ms in duration was delivered between superior vena cava and right ventricular catheter electrodes 10 s after the electrical induction of ventricular fibrillation. Six dogs underwent an up/down defibrillation protocol randomized with or without the patch on the heart. Results. Mean 50% (+/-SD) probability point for energy doubled with the conductive patch on the heart, from 8.0 +/- 3.2 to 16.8 +/- 7.0 J (p < 0.01), and leading edge voltage increased from 334 +/- 64 to 477 +/- 98 V (p < 0.01). Mean 50% probability points for energy and leading-edge voltage were not significantly changed when the procedure was repeated using a nonconductive patch in another six dogs as a control group. In a saline-saturated foam model, measurements from electrodes plated around and under the patch revealed a 72% mean decrease in the potential gradient in the foam under the conductive patch. Conclusions. A passive defibrillator patch can markedly increase the energy requirements for defibrillation, probably by decreasing the potential gradient under the patch. These results suggest the use of caution when passive electrodes are present, for example, when a patient receives a nonthoracotomy defibrillator system while epicardial electrodes from a previously implanted system are left in place.
引用
收藏
页码:1373 / 1379
页数:7
相关论文
共 31 条
[1]  
BOURLAND JD, 1978, MED INSTRUM, V12, P38
[2]   DOUBLE AND TRIPLE SEQUENTIAL SHOCKS REDUCE VENTRICULAR DEFIBRILLATION THRESHOLD IN DOGS WITH AND WITHOUT MYOCARDIAL-INFARCTION [J].
CHANG, MS ;
INOUE, H ;
KALLOK, MJ ;
ZIPES, DP .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1986, 8 (06) :1393-1405
[3]   THE POTENTIAL GRADIENT FIELD CREATED BY EPICARDIAL DEFIBRILLATION ELECTRODES IN DOGS [J].
CHEN, PS ;
WOLF, PD ;
CLAYDON, FJ ;
DIXON, EG ;
VIDAILLET, HJ ;
DANIELEY, ND ;
PILKINGTON, TC ;
IDEKER, RE .
CIRCULATION, 1986, 74 (03) :626-636
[4]   COMPARISON OF THE DEFIBRILLATION THRESHOLD AND THE UPPER LIMIT OF VENTRICULAR VULNERABILITY [J].
CHEN, PS ;
SHIBATA, N ;
DIXON, EG ;
MARTIN, RO ;
IDEKER, RE .
CIRCULATION, 1986, 73 (05) :1022-1028
[5]  
COX JL, 1989, CIRCULATION, V79, P163
[6]   THE RELATIONSHIP BETWEEN SUCCESSFUL DEFIBRILLATION AND DELIVERED ENERGY IN OPEN-CHEST DOGS - REAPPRAISAL OF THE DEFIBRILLATION THRESHOLD CONCEPT [J].
DAVY, JM ;
FAIN, ES ;
DORIAN, P ;
WINKLE, RA .
AMERICAN HEART JOURNAL, 1987, 113 (01) :77-84
[7]   SEQUENTIAL PULSE INTERNAL DEFIBRILLATION - IS THERE AN ADVANTAGE TO SWITCHED CURRENT PATHWAYS [J].
FAIN, ES ;
SWEENEY, MB ;
FRANZ, MR .
AMERICAN HEART JOURNAL, 1989, 118 (04) :717-724
[8]   OPTIMUM ELECTROLYTIC CHLORIDING OF SILVER ELECTRODES [J].
GEDDES, LA ;
BAKER, LE ;
MOORE, AG .
MEDICAL & BIOLOGICAL ENGINEERING, 1969, 7 (01) :49-&
[9]   FUNDAMENTAL CRITERIA UNDERLYING THE EFFICACY AND SAFETY OF DEFIBRILLATING CURRENT WAVEFORMS [J].
GEDDES, LA ;
NIEBAUER, MJ ;
BABBS, CF ;
BOURLAND, JD .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1985, 23 (02) :122-130
[10]   CONTOUR GRAPH FOR RELATING PER CENT SUCCESS IN ACHIEVING VENTRICULAR DEFIBRILLATION TO DURATION, CURRENT, AND ENERGY CONTENT OF SHOCK [J].
GOLD, JH ;
SCHUDER, JC ;
STOECKLE, H .
AMERICAN HEART JOURNAL, 1979, 98 (02) :207-212