NONINVASIVE TEMPERATURE IMAGING USING DIFFUSION MRI

被引:85
作者
DELANNOY, J
CHEN, CN
TURNER, R
LEVIN, RL
LEBIHAN, D
机构
[1] NIH,WARREN GRANT MAGNUSON CLIN CTR,DEPT DIAGNOST RADIOL,BETHESDA,MD 20892
[2] NIH,DIV RES SERV,BIOMED ENGN & INSTRUMENTAT PROGRAM,BETHESDA,MD 20892
[3] NIH,RADIAT ONCOL BRANCH,COP,DCT,BETHESDA,MD 20892
[4] NCI,BETHESDA,MD 20892
关键词
D O I
10.1002/mrm.1910190224
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Efficacy and safety considerations for cancer therapy with hyperthermia require accurate temperature measurements throughout the heated volume. We report the use of molecular diffusion, whose temperature dependence is well known. A dedicated hyperthermia applicator was built, combining a MRI gradient coil and a rf coil. Diffusion and derived temperature images were obtained with a 1 × 2 mm pixel size on a polyacrylamide gel phantom using a clinical 1.5‐T whole body MRI system. Temperatures determined from these images using 1 cm2 regions of interest were found to be within 0.2° of those recorded from the thermocouples and fiber‐optic probes placed inside the gel. © 1991 Academic Press, Inc. Copyright © 1991 Wiley‐Liss, Inc., A Wiley Company
引用
收藏
页码:333 / 339
页数:7
相关论文
共 15 条
  • [1] THE EFFECTS OF RANDOM DIRECTIONAL DISTRIBUTED FLOW IN NUCLEAR-MAGNETIC-RESONANCE IMAGING
    AHN, CB
    LEE, SY
    NALCIOGLU, O
    CHO, ZH
    [J]. MEDICAL PHYSICS, 1987, 14 (01) : 43 - 48
  • [2] EFFICIENT DECOUPLER COIL DESIGN WHICH REDUCES HEATING IN CONDUCTIVE SAMPLES IN SUPERCONDUCTING SPECTROMETERS
    ALDERMAN, DW
    GRANT, DM
    [J]. JOURNAL OF MAGNETIC RESONANCE, 1979, 36 (03) : 447 - 451
  • [3] CETAS TC, 1984, CANCER RES, V44, P4805
  • [4] DICKINSON RJ, 1986, J COMPUT ASSIST TOMO, V10, P468
  • [5] GIBBS FA, 1984, HYPERTHERMIC ONCOLOG, V2, P155
  • [6] EXPERIMENTAL CHARACTERIZATION OF THE MINIANNULAR PHASED-ARRAY AS A HYPERTHERMIA APPLICATOR
    GUERQUIN-KERN, JL
    HAGMANN, MJ
    LEVIN, RL
    [J]. MEDICAL PHYSICS, 1987, 14 (04) : 674 - 680
  • [7] HAHN GM, 1982, HYPERTHERMIA CANCER
  • [8] Knuttel B, 1986, Recent Results Cancer Res, V101, P109
  • [9] MR IMAGING OF INTRAVOXEL INCOHERENT MOTIONS - APPLICATION TO DIFFUSION AND PERFUSION IN NEUROLOGIC DISORDERS
    LEBIHAN, D
    BRETON, E
    LALLEMAND, D
    GRENIER, P
    CABANIS, E
    LAVALJEANTET, M
    [J]. RADIOLOGY, 1986, 161 (02) : 401 - 407
  • [10] SEPARATION OF DIFFUSION AND PERFUSION IN INTRAVOXEL INCOHERENT MOTION MR IMAGING
    LEBIHAN, D
    BRETON, E
    LALLEMAND, D
    AUBIN, ML
    VIGNAUD, J
    LAVALJEANTET, M