4-YEAR EXPERIENCE WITH THE CARBOMEDICS VALVE - THE NORTH-AMERICAN EXPERIENCE

被引:41
作者
COPELAND, JG [1 ]
SETHI, GK [1 ]
BASSETT, JS [1 ]
BROWN, G [1 ]
CRAVER, JM [1 ]
GAGO, O [1 ]
JAMIESON, WRE [1 ]
MICHLER, RE [1 ]
MCENANY, T [1 ]
PELLETIER, LC [1 ]
REED, GF [1 ]
WEINTRAUB, RM [1 ]
ZAPOLANSKI, A [1 ]
机构
[1] UNIV ARIZONA,HLTH SCI CTR,TUCSON,AZ 85724
关键词
D O I
10.1016/0003-4975(94)90719-6
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A multicenter prospective trial of the CarboMedics valve was initiated in July 1988. Twelve North American centers were included for the purpose of establishing the safety and efficacy of this new rotatable, bileaflet, pyrolytic-Carbon prosthesis. The study included 786 patients (using 891 valves) enrolled between July 1988 and August 1992. There were 447 male and 339 female patients from ages 2 to 83 years. Of the 618 primary replacement operations (including double valves), 393 were aortic and 292 were mitral valve replacements. Mean follow-up time was 16.1 months; 94.7% of patients had completed documentation up to and including one year. New York Heart Association functional class was III or IV for 84.4% of patients preoperatively, and I and II in over 96% of patients for the first three postoperative years. Survival for the entire group was 93% (622) at 1 month, 89% (504) at 1 year, 87% (213) at 2 years, and 85% (41) at 3 years. There were no significant survival differences between primary and repeat replacements or between aortic, mitral, and double-valve replacements. Primary replacement patients with concomitant procedures (n = 241) had slightly decreased survival rates compared with those who underwent isolated procedures (isolated survival rate, 92% [98]; concomitant procedure survival rate, 81% [59] at 2 years; p = 0.002). There was no valve failure. Linearized rates (events per 100 patient-years) in primary replacement patients after 30 days included the following: thrombosis, 0.39; thromboembolism, 0.92; hemolysis, 0.53; anticoagulant-related hemorrhage, 2.36; paravalvular leak, 0.92; endocarditis, 0.53; reoperation, 0.92; and explantation, 0.66. We conclude that this valve, which recently has been recommended for approval by the Food and Drug Administration advisory panel, is safe and efficacious for use.
引用
收藏
页码:630 / 638
页数:9
相关论文
共 17 条
[1]   10 YEARS EXPERIENCE WITH THE ST JUDE MEDICAL VALVE PROSTHESIS [J].
AROM, KV ;
NICOLOFF, DM ;
KERSTEN, TE ;
NORTHRUP, WF ;
LINDSAY, WG ;
EMERY, RW .
ANNALS OF THORACIC SURGERY, 1989, 47 (06) :831-837
[2]   DEATH AND OTHER TIME-RELATED EVENTS AFTER VALVE-REPLACEMENT [J].
BLACKSTONE, EH ;
KIRKLIN, JW .
CIRCULATION, 1985, 72 (04) :753-767
[3]   COMPARATIVE-STUDY OF THE HYDRODYNAMIC FUNCTION OF THE CARBOMEDICS VALVE [J].
BUTTERFIELD, M ;
FISHER, J ;
DAVIES, GA ;
SPYT, TJ .
ANNALS OF THORACIC SURGERY, 1991, 52 (04) :815-820
[4]   ECHOCARDIOGRAPHIC DESCRIPTION OF THE CARBOMEDICS BILEAFLET PROSTHETIC HEART-VALVE [J].
CHAMBERS, J ;
CROSS, J ;
DEVERALL, P ;
SOWTON, E .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (02) :398-405
[5]   THE ST-JUDE VALVE - ANALYSIS OF THROMBOEMBOLISM, WARFARIN-RELATED HEMORRHAGE, AND SURVIVAL [J].
CZER, LSC ;
MATLOFF, JM ;
CHAUX, A ;
DEROBERTIS, M ;
STEWART, ME ;
GRAY, RJ .
AMERICAN HEART JOURNAL, 1987, 114 (02) :389-397
[6]  
DELUCA L, 1993, J THORAC CARDIOV SUR, V106, P1158
[7]   COMPUTATIONAL FLUID-DYNAMICS OF ARTIFICIAL-HEART VALVES [J].
DUBINI, G ;
PIETRABISSA, R ;
FUMERO, R .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1991, 14 (06) :338-342
[8]   GUIDELINES FOR REPORTING MORBIDITY AND MORTALITY AFTER CARDIAC VALVULAR OPERATIONS [J].
EDMUNDS, LH ;
CLARK, RE ;
COHN, LH ;
MILLER, C ;
WEISEL, RD .
ANNALS OF THORACIC SURGERY, 1988, 46 (03) :257-259
[9]   THROMBOTIC AND BLEEDING COMPLICATIONS OF PROSTHETIC HEART-VALVES [J].
EDMUNDS, LH .
ANNALS OF THORACIC SURGERY, 1987, 44 (04) :430-445
[10]  
GEIRAN O, 1990, SURGERY HEART VALVE, P551