LARGE-VOLUME INJECTIONS IN GAS-CHROMATOGRAPHY ATOMIC-EMISSION DETECTION - AN APPROACH FOR TRACE-LEVEL DETECTION IN WATER ANALYSIS

被引:33
作者
RINKEMA, FD [1 ]
LOUTER, AJH [1 ]
BRINKMAN, UAT [1 ]
机构
[1] FREE UNIV AMSTERDAM, DEPT ANALYT CHEM, 1081 HV AMSTERDAM, NETHERLANDS
关键词
D O I
10.1016/0021-9673(94)80476-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The ruggedness and analytical performance of on-line capillary gas chromatography-atomic emission detection (GC-AED) have been studied using 100-mu l injections of sample solutions in ethyl acetate, via a loop-type interface. A series of organophosphorus compounds were selected as test analytes; they were monitored using the carbon, sulphur, nitrogen, chlorine, bromine and phosphorus channels. The system showed no flame-outs or other maintenance problems even after 300 large-volume injections. The analytical potential of the system, expressed in terms of repeatability, linearity and minimum detectable amount, was not affected and a 100-fold increase in analyte detectability, in terms of concentration units, compared with a conventional 1-mu l injection was observed. As an application, GC-AED was combined off-line with solid-phase extraction. Several environmental contaminants were preconcentrated from river and tap water samples, and 20% (100 mu l) of the ethyl acetate eluent were directly analysed. With a sample volume of only 10 ml, the detection limits of the organophosphorus pesticides typically were ca. 0.1 mu g/l.
引用
收藏
页码:289 / 297
页数:9
相关论文
共 15 条
  • [1] BULTERMAN AJ, 1993, HRC-J HIGH RES CHROM, V16, P397, DOI 10.1002/jhrc.1240160703
  • [2] LC GC, SFC GC, AND SFE GC - INTERFACING
    DAVIES, IL
    RAYNOR, MW
    KITHINJI, JP
    BARTLE, KD
    WILLIAMS, PT
    ANDREWS, GE
    [J]. ANALYTICAL CHEMISTRY, 1988, 60 (11) : A683 - &
  • [3] GOOSENS EC, UNPUB
  • [4] HYPHENATED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY CAPILLARY GAS-CHROMATOGRAPHY
    GROB, K
    [J]. JOURNAL OF CHROMATOGRAPHY, 1992, 626 (01): : 25 - 32
  • [5] DETERMINATION OF ORGANOPHOSPHORUS PESTICIDES IN AQUEOUS SAMPLES BY ONLINE MEMBRANE DISK EXTRACTION AND CAPILLARY GAS-CHROMATOGRAPHY
    KWAKMAN, PJM
    VREULS, JJ
    BRINKMAN, UAT
    GHIJSEN, RT
    [J]. CHROMATOGRAPHIA, 1992, 34 (1-2) : 41 - 47
  • [6] FULLY AUTOMATED WATER ANALYZER BASED ON ONLINE SOLID-PHASE EXTRACTION GAS-CHROMATOGRAPHY
    LOUTER, AJH
    BRINKMAN, UAT
    GHIJSEN, RT
    [J]. JOURNAL OF MICROCOLUMN SEPARATIONS, 1993, 5 (04) : 303 - 315
  • [7] LOUTER AJH, 1994, IN PRESS INT J ENV A, V54
  • [8] DETERMINATION OF TRIAZINES AND ORGANOPHOSPHORUS PESTICIDES IN WATER SAMPLES USING SOLID-PHASE EXTRACTION
    MOLTO, JC
    PICO, Y
    FONT, G
    MANES, J
    [J]. JOURNAL OF CHROMATOGRAPHY, 1991, 555 (1-2): : 137 - 145
  • [9] UDEN PC, 1972, ACS SYM SER, V479, P11
  • [10] ISOLATION AND CONCENTRATION OF ORGANO-PHOSPHORUS PESTICIDES FROM WATER USING A C-18 REVERSED PHASE
    VINUESA, JM
    CORTES, JCM
    CANAS, CI
    PEREZ, GF
    [J]. JOURNAL OF CHROMATOGRAPHY, 1989, 472 (02): : 365 - 370