MEASUREMENT ACCURACY - COMPARISON OF 2 INTRAORAL DIGITAL RADIOGRAPHIC SYSTEMS, RADIOVISIOGRAPHY-S AND FLASHDENT, WITH ANALOG FILM

被引:6
作者
SCARFE, WC
NORTON, S
FARMAN, AG
机构
[1] The University of Louisville School of Dentistry, KY
关键词
RADIOGRAPHY; DENTAL; IMAGE-ENHANCEMENT; COMPUTER-ASSISTED;
D O I
10.1259/dmfr.24.4.9161164
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To compare the accuracy of measurement algorithms incorporated into the software provided with two CCD-based digital radiographic systems and compare assessments using intra-oral film. Methods. A test object with three radiopaque reference points was imaged using Ektaspeed intra-oral film, both with and without an overlaid 1 mm reference grid, the RVG-S, the measurement software of which overlays a 2 mm grid, and the FlashDent, which uses a mouse-driven cursor to estimate distances. Geometric factors, including source-to-object and receptor-to-object distances, vertical and horizontal cone angulation, and vertical angulation of receptor, were successively altered. Six viewers estimated vertical, horizontal and diagonal distances for each image. Results. Increased source-to-object distance resulted in a 5-10% reduction in image dimensions, whereas increased receptor-to-object distance produced a magnification of 15-20% with all techniques. Increased vertical and horizontal cone angulation resulted in slight increases in magnification (0-5%) whereas greater receptor angulation resulted in noticeable magnification under all test conditions (10-15%). Estimated values for vertical, horizontal and diagonal measurements were approximately 5-10% higher than the actual dimensions, irrespective of the technique used. The FlashDent mouse-driven cursor system and direct measurement of distances from conventional film were more accurate in horizontal and diagonal dimension estimates than using an overlaid grid. Intra-observer and interobserver differences were found for specific situations. Conclusion. The mouse-driven computerized measurement cursor and intra-oral film assessment using a ruler provided the most accurate, reliable and consistent dimensional measurements.
引用
收藏
页码:215 / 220
页数:6
相关论文
empty
未找到相关数据