EFFECT OF REFRACTION ON OPTICAL MICROSCOPIC MEASUREMENT OF INTERNAL BLOOD-VESSEL DIAMETER AND ITS CORRECTION

被引:3
作者
CHEN, EQ [1 ]
LAM, CF [1 ]
PERIASAMY, A [1 ]
机构
[1] MED UNIV S CAROLINA,DEPT ANAT,CHARLESTON,SC 29425
来源
JOURNAL OF MICROSCOPY-OXFORD | 1991年 / 164卷
关键词
OPTICAL MICROSCOPE; BLOOD VESSEL; INTERNAL DIAMETER; REFRACTION; MEASUREMENT; CORRECTION;
D O I
10.1111/j.1365-2818.1991.tb03210.x
中图分类号
TH742 [显微镜];
学科分类号
摘要
In optical microscopic measurement of internal blood-vessel diameters, the effect of refraction must be taken into account to ensure accuracy of the result. This effect is discussed and an analytical correction formula derived. Phantom blood vessels with known internal and external diameters were used to test the validity of the correction formula. The errors obtained prior to correction were reduced significantly after correction.