FLOW-THROUGH ANODIC-STRIPPING COULOMETRY AND ANODIC-STRIPPING COULOMETRY WITH COLLECTION FOR THE SIMULTANEOUS ABSOLUTE DETERMINATION OF COPPER, LEAD, CADMIUM AND ZINC

被引:13
作者
BEINROHR, E
TSCHOPEL, P
TOLG, G
NEMETH, M
机构
[1] MAX PLANCK INST MET RES, REINSTSTOFFANALYT LAB, W-4600 DORTMUND 1, GERMANY
[2] INST SPEKTROCHEM & ANGEW SPEKT, W-4600 DORTMUND 1, GERMANY
关键词
COULOMETRY; FLOW SYSTEM; STRIPPING VOLTAMMETRY; CADMIUM; COPPER; LEAD; ZINC;
D O I
10.1016/0003-2670(93)80140-G
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Flow-through electrochemical cells with porous working electrodes made of crushed reticulated vitreous carbon and plated with mercury were used for the absolute determination of copper, lead, cadmium and zinc by anodic stripping coulometry and anodic stripping coulometry with collection in a flow system. The system consisted of a peristaltic pump, pneumatic pulse damper, electrochemical flow-through scrubber to remove impurities and dissolved oxygen from the flowing electrolyte, sample injection port and two serial flow-through cells with porous working electrodes. Trace metals from the injected sample volume were deposited at the working electrode of the first cell, then stripped by scanning the potential to positive values and finally collected at the second cell. Both the stripping and collection currents were registered and integrated. The conditions for a quantitative deposition, Stripping and collection were investigated.
引用
收藏
页码:13 / 25
页数:13
相关论文
共 27 条
[1]  
Bard A. J., 1980, ELECTROCHEMICAL METH, V2nd, P406
[2]   DESIGN AND CHARACTERIZATION OF FLOW-THROUGH COULOMETRIC CELLS WITH POROUS WORKING ELECTRODES MADE OF CRUSHED VITREOUS CARBON [J].
BEINROHR, E ;
NEMETH, M ;
TSCHOPEL, P ;
TOLG, G .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 343 (07) :566-575
[3]  
BEINROHR E, IN PRESS FRESENIUS J
[4]   ANODIC-STRIPPING VOLTAMMETRY AT A RETICULATED MERCURY VITREOUS CARBON ELECTRODE [J].
BLAEDEL, WJ ;
WANG, J .
ANALYTICAL CHEMISTRY, 1979, 51 (11) :1724-1728
[5]   FLOW ELECTROLYSIS ON A RETICULATED VITREOUS CARBON ELECTRODE [J].
BLAEDEL, WJ ;
WANG, J .
ANALYTICAL CHEMISTRY, 1979, 51 (07) :799-802
[6]   ELECTROCHEMICAL DETECTOR BASED ON A RETICULATED VITREOUS CARBON WORKING ELECTRODE FOR LIQUID-CHROMATOGRAPHY AND FLOW-INJECTION ANALYSIS [J].
CURRAN, DJ ;
TOUGAS, TP .
ANALYTICAL CHEMISTRY, 1984, 56 (04) :672-678
[7]   LINEAR POTENTIAL-SWEEP VOLTAMMETRY AT A PLANE MERCURY-FILM ELECTRODE [J].
DEVRIES, WT ;
VANDALEN, E .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1967, 14 (03) :315-&
[8]   ANODIC STRIPPING VOLTAMMETRY WITH A GLASSY CARBON ELECTRODE MERCURY-PLATED IN-SITU [J].
FLORENCE, TM .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1970, 27 (02) :273-&
[9]   DIRECT REDUCTIVE AMPEROMETRIC DETERMINATION OF NITRATE AT A COPPER ELECTRODE FORMED INSITU IN A CAPILLARY-FILL SENSOR DEVICE [J].
FOGG, AG ;
SCULLION, SP ;
EDMONDS, TE ;
BIRCH, BJ .
ANALYST, 1991, 116 (06) :573-579
[10]   STRIPPING VOLTAMMETRY WITH COLLECTION AT A ROTATING-RING-DISK ELECTRODE [J].
JOHNSON, DC ;
ALLEN, RE .
TALANTA, 1973, 20 (03) :305-313