ONLINE MONITORING OF THE DELIVERY OF THE HEMODIALYSIS PRESCRIPTION

被引:43
作者
KESHAVIAH, PR
EBBEN, JP
EMERSON, PF
机构
[1] Baxter Clinical Engineering, Baxter Healthcare Corporation, Minneapolis, 55 404, MN
[2] Baxter Clinical Engineering, Minneapolis Medical Research Foundation, Minneapolis, Minnesota
关键词
HEMODIALYSIS PRESCRIPTION; ONLINE MONITORING;
D O I
10.1007/BF00867675
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
The BioStat 1000 is a new device which employs dialysate-based urea kinetics to calculate the dose of-dialysis (Kt/V) based on a two-pool model and protein catabolic rate (PCR). Previous methods relying on blood sampling techniques were subject to error and difficult to implement. This paper describes the features of the Biostat and the results of the first clinical validation study with an early prototype. The BioStat was found to compare favorably with the reference method of direct dialysate quantification (mDDQ) which had been modified to obtain a ''two-pool'' Kt/V. In 31 patients no significant difference was found between mean Kt/V from the mDDQ and the mean Kt/V from the BioStat (1.35 +/- 0.33 versus 1.38 +/- 0.36, respectively). The PCR was also not significantly different (53.4 +/- 18.5 g/day versus 51.8 +/- 16 g/day, respectively). The BioStat was demonstrated to be a convenient method producing reliable results.
引用
收藏
页码:S2 / S8
页数:7
相关论文
共 23 条
[1]  
Farrell P.C., Dialysis kinetics, ASAIO Primers in Artificial Organs, 4, pp. 1-30, (1988)
[2]  
Depner T.A., Prescribing hemodialysis: a guide to urea modeling, (1991)
[3]  
Gotch F., Kinetic modeling in hemodialysis, Clinical dialysis, pp. 118-146, (1990)
[4]  
Sargent J.A., Gotch F.A., Principles and biophysics of dialysis, Replacement of renal function by dialysis, pp. 87-143, (1989)
[5]  
Ilstrup K., Hanson G., Shapiro W., Keshaviah P., Examining the foundations of urea kinetics, Trans Am Soc Artif Intern Organs, 31, pp. 164-168, (1985)
[6]  
Daugirdas J.T., Second generation logarithmic estimates of single-pool variable volume Kt/V: an analysis of Error, J Am Soc Nephrol, 4, pp. 1205-1213, (1993)
[7]  
Barth R., Direct calculation of KT/V: a simplified approach to monitoring of hemodialysis, Nephron, 50, pp. 191-195, (1988)
[8]  
Frost T., Kerr D., Kinetics of hemodialysis: a theoretical study of the removal of solutes in chronic renal failure compared to normal health, Kidney Int, 12, pp. 41-50, (1977)
[9]  
Delmez J.A., Windus D.W., Hemodialysis prescription and delivery in a metropolitan community. The St. Louis Nephrology Study Group, Kidney Int, 41, pp. 1023-1028, (1992)
[10]  
Held P., Levin N., Bovbjerg R., Pauly M.V., Diamond L.H., Mortality and duration of hemodialysis treatment, JAMA, 265, pp. 871-875, (1991)