A 5-CHANNEL, DIAMOND PHOTOCONDUCTING X-RAY-DETECTOR ARRAY FOR Z-PINCH EXPERIMENTS

被引:36
作者
SPIELMAN, RB
机构
关键词
D O I
10.1063/1.1143490
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have built a five-channel, x-ray detector array based on diamond photoconducting detectors (PCDs). The diamond elements have dimensions of 3 mm x 1 mm X 1 mm (or 0.5 mm). We use diamond PCDs for their stability, flat spectral response, and low leakage currents. The good time response of diamond PCDs is due to the 100-ps electron/hole recombination time. Filters were designed to give information in the 1-10-keV spectral region. Calibration of the diamond PCDs showed sensitivities between 4 and 7 x 10(-4) A/W for a bias of 100 V. We shall present data from z-pinch experiments on Saturn.
引用
收藏
页码:5056 / 5058
页数:3
相关论文
共 10 条
[1]  
BLOOMQUIST DD, 1987, 6TH P IEEE PULS POW, P310
[2]  
HANSON DL, 1980, B AM PHYS SOC, V25, P890
[3]  
JONES LA, 1986, J APPL PHYS, V58, P1711
[4]   SOFT-X-RAY DETECTION WITH DIAMOND PHOTOCONDUCTIVE DETECTORS [J].
KANIA, DR ;
PAN, L ;
KORNBLUM, H ;
BELL, P ;
LANDEN, ON ;
PIANETTA, P .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (10) :2765-2767
[5]   FLAT RESPONSE DETECTORS FOR THE VACUUM ULTRAVIOLET AND SOFT-X-RAY REGION - INP-FE PHOTOCONDUCTORS [J].
KANIA, DR ;
IVERSON, AE ;
SMITH, DL ;
WAGNER, RS ;
HAMMOND, RB ;
STETLER, KA .
JOURNAL OF APPLIED PHYSICS, 1986, 60 (07) :2596-2599
[6]  
KEUTCH TF, 1982, APPL PHYS LETT, V41, P403
[7]   Z-PINCH OF A GAS-JET [J].
SHILOH, J ;
FISHER, A ;
ROSTOKER, N .
PHYSICAL REVIEW LETTERS, 1978, 40 (08) :515-518
[8]  
SPIELMAN RB, 1989, AIP CONF PROC, V195, P3, DOI 10.1063/1.38844
[9]   PHOTOCONDUCTING X-RAY-DETECTORS FOR Z-PINCH EXPERIMENTS [J].
SPIELMAN, RB ;
HSING, WW ;
HANSON, DL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (08) :1804-1806
[10]  