FLOW CELL WITH DOUBLE SLOPE FACTOR FOR POTENTIOMETRIC DETERMINATION OF FLUORIDE AT LOW CONCENTRATIONS

被引:11
作者
BORZITSKY, JA [1 ]
DVININ, A [1 ]
PETRUKHIN, OM [1 ]
URUSOV, YI [1 ]
机构
[1] DI MENDELEEV MOSCOW CHEM TECHNOL INST,MOSCOW 125820,RUSSIA
关键词
FLOW ANALYSIS; ION-SELECTIVE ELECTRODE; SENSITIVITY; FLUORIDE;
D O I
10.1039/an9931800859
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple and reliable double-cell device with three ion-selective membranes for fluoride determinations in a continuous-flow system with an increased slope factor is presented. The relative standard deviation for the analysis of environmental waters was 0.01-0.03 and the slope factor was 113-44 mV pF-1, depending on the fluoride concentration level; the sampling rate was 120 h-1.
引用
收藏
页码:859 / 861
页数:3