ISOTOPE-DILUTION GLOW-DISCHARGE MASS-SPECTROMETRY AS APPLIED TO THE ANALYSIS OF LEAD IN WASTE OIL SAMPLES

被引:9
作者
BARSHICK, CM [1 ]
SMITH, DH [1 ]
WADE, JW [1 ]
BAYNE, CK [1 ]
机构
[1] OAK RIDGE NATL LAB,DIV COMP & TELECOMMUN,OAK RIDGE,TN 37831
关键词
GLOW DISCHARGE MASS SPECTROMETRY; SOLUTION RESIDUE ANALYSIS; ISOTOPE DILUTION; INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECTROMETRY;
D O I
10.1039/ja9940900083
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The application of isotope dilution (ID) to glow discharge mass spectrometry (GDMS) has been evaluated. Initial work with the analysis of solution residues has been modified to accommodate introduction of an isotopic spike. In evaluating ID-GDMS, elemental lead concentrations were determined in oil leachate samples. After equilibration of the sample and spike, a few hundred microliter volume of solution was mixed with a conducting host matrix. This slurry was then dried, homogenized and pressed into a GD cathode; after electrode formation, routine GDMS analysis (isotopic) was performed. Internal precision (within sample variation) was estimated to be better than 5% relative standard deviation (RSD) (1sigma) through the use of a computer model of the ID equation. External precision (between-sample variation) was better than 2% RSD (1sigma), a substantial improvement over most conventional GDMS analyses. For comparison purposes (i.e., where multiple isotopic spikes are impractical), a calibration curve was applied to the same data and gave external precision of better than 15% RSD. Both calibration curve and ID results compare well with those obtained by inductively coupled plasma atomic emission spectrometry.
引用
收藏
页码:83 / 87
页数:5
相关论文
共 17 条
[1]   DETERMINATION OF PICOGRAM AMOUNTS OF TC-99 BY RESIN BEAD MASS-SPECTROMETRIC ISOTOPE-DILUTION [J].
ANDERSON, TJ ;
WALKER, RL .
ANALYTICAL CHEMISTRY, 1980, 52 (04) :709-713
[2]   ANALYSIS OF SOLUTION RESIDUES BY GLOW-DISCHARGE MASS-SPECTROMETRY [J].
BARSHICK, CM ;
DUCKWORTH, DC ;
SMITH, DH .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1993, 4 (01) :47-53
[3]  
BARSHICK CM, UNPUB ANAL CHEM
[4]  
BAYNE CK, 1991, ORNLCSDTM282 REP
[5]  
DEBIEVRE P, 1978, ADV MASS SPECTROM, V7, P395
[6]  
FARRAR H, 1972, TRACE ANAL MASS SPEC, P239
[7]   STABLE ISOTOPE-DILUTION ANALYSIS OF HYDROLOGIC SAMPLES BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
GARBARINO, JR ;
TAYLOR, HE .
ANALYTICAL CHEMISTRY, 1987, 59 (11) :1568-1575
[8]   APPLICATION OF AN ECONOMICAL AND SMALL THERMAL IONIZATION MASS-SPECTROMETER FOR ACCURATE ANION TRACE ANALYSES [J].
HEUMANN, KG ;
SCHINDLMEIER, W ;
ZEININGER, H ;
SCHMIDT, M .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1985, 320 (05) :457-462
[9]   HIGH-ACCURACY IN THE ELEMENT ANALYSIS BY MASS-SPECTROMETRY [J].
HEUMANN, KG .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1986, 324 (06) :601-611
[10]   TRACE DETERMINATION AND ISOTOPIC ANALYSIS OF THE ELEMENTS IN LIFE SCIENCES BY MASS-SPECTROMETRY [J].
HEUMANN, KG .
BIOMEDICAL MASS SPECTROMETRY, 1985, 12 (09) :477-488