MAGNETIZATION PREPARED RAPID GRADIENT-ECHO (MP-RAGE) MR IMAGING OF THE LIVER - COMPARISON WITH SPIN-ECHO IMAGING

被引:35
作者
DELANGE, EE
MUGLER, JP
BERTOLINA, JA
GAY, SB
JANUS, CL
BROOKEMAN, JR
机构
[1] Department of Radiology, University of Virginia Health Sciences Center, Charlottesville
关键词
MAGNETIC RESONANCE (MR); TECHNOLOGY; PULSE SEQUENCES; RAPID IMAGING; LIVER NEOPLASMS; MR STUDIES; LIVER;
D O I
10.1016/0730-725X(91)90031-G
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We have implemented an MR technique that employs a rapid gradient echo sequence, preceded by magnetization preparation pulses to provide T1- and T2-weighted tissue contrast. With this technique, which can be identified as a member of a new family of pulse sequences, generically named Magnetization Prepared RApid Gradient Echo (MP-RAGE), very short repetition times are used, allowing acquisition times of less than one second and images virtually free of motion-induced artifacts during quiet respiration. Fifteen patients with known liver lesions (metastases, hemangiomas, and cysts) were examined using T1- and T2-weighted 2-dimensional MP-RAGE sequences, and the images were compared with conventional T1- and multi-echo T2-weighted spin-echo (SE) sequences. Signal difference-to-noise ratios (SD/Ns) of the lesions were calculated for all pulse sequences using corresponding axial images and were normalized for voxel volume. The mean normalized SD/Ns of the MP-RAGE sequences were generally comparable to those for the SE sequences. In addition, there were no noticeable respiratory artifacts on the MP-RAGE images whereas these were clearly present on the T2-weighted SE images and to a lesser degree on the T1-weighted SE images. It is concluded that the MP-RAGE technique could become an important method for evaluating the liver for focal disease.
引用
收藏
页码:469 / 476
页数:8
相关论文
共 22 条
  • [1] Haase, Matthaei, Bartkowski, Duhmke, Leibfritz, Inversion recovery snapshot FLASH MR imaging, J. Comput. Assist. Tomogr., 13, pp. 1036-1040, (1989)
  • [2] Haase, Snapshot FLASH MRI. Applications to T<sub>1</sub>, T<sub>2</sub>, and chemical-shift imaging, Magn. Reson. Med., 13, pp. 77-89, (1990)
  • [3] Unger, Cohen, Gatenby, Clair, Brown, Nelson, McGlone, Single breath-holding scans of the abdomen using FISP and FLASH at 1.5T, J. Comput. Assist. Tomogr., 12, pp. 575-583, (1988)
  • [4] Frahm, Haase, Matthaei, Rapid NMR imaging of dynamic processes using the FLASH technique, Magn. Reson. Med., 3, pp. 321-327, (1986)
  • [5] Kiefer, Deimling, Finelli, Ultrafast measurement of T<sub>1</sub>- and T<sub>2</sub>-weighted images with “Snapshot”-FLASH, Book of Abstracts, 1, (1989)
  • [6] Conturo, Kessler, Beth, Cooperative T<sub>1</sub> and T<sub>2</sub> effects on contrast using a new driven inversion spin-echo (DISE) MRI pulse sequence, Magn. Reson. Med., 15, pp. 397-419, (1990)
  • [7] Mugler, Brookeman, Optimization of pulse sequence parameters in “Snapshot-FLASH” imaging, Magn. Reson. Imaging, 8, 1 S, (1990)
  • [8] Glantz, Primer of Biostatistics, (1987)
  • [9] Reinig, Dwyer, Miller, Frank, Adams, Chang, Liver metastases: Detection with MR imaging at 0.5 and 1.5 T, Radiology, 170, pp. 149-153, (1989)
  • [10] Foley, Kneeland, Cates, Kellman, Lawson, Middleton, Hendrick, Contrast optimization for the detection of focal hepatic lesions by MR imaging at 1.5 T, AJR, 149, pp. 1155-1160, (1987)