Transdermal penetration studies ''in vitro'' are tedious and time consuming, and sometimes require complex sampling schedules. If automatic equipment is used, full validation is mandatory. The aim of this study was to validate an automated ''in vitro'' sampling system using modified Franz diffusion cells. The reproductibility of sample and replace volumes was assessed. The dilution effect induced by the replaced volume was also studied and some equations in order to correct this factor were propossed. Ketorolac trometamol has been used as a model drug in this study.
引用
收藏
页码:2267 / 2283
页数:17
相关论文
共 4 条
[1]   AN AUTOMATED-METHOD FOR CONTINUOUSLY MONITORING DIFFUSION CELLS IN SKIN PENETRATION STUDIES [J].
AKAZAWA, M ;
ITOH, T ;
MASAKI, K ;
NGHIEM, BT ;
TSUZUKI, N ;
KONISHI, R ;
HIGUCHI, T .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1989, 50 (01) :53-60
[2]   INVESTIGATIONS INTO THE PHARMACODYNAMIC EFFECTS OF DERMALLY ADMINISTERED MICROEMULSIONS CONTAINING BETA-BLOCKERS [J].
KEMKEN, J ;
ZIEGLER, A ;
MULLER, BW .
JOURNAL OF PHARMACY AND PHARMACOLOGY, 1991, 43 (10) :679-684
[3]   A NEW DIFFUSION CELL - AN AUTOMATED-METHOD FOR MEASURING THE PHARMACEUTICAL AVAILABILITY OF TOPICAL DOSAGE FORMS [J].
MARTIN, B ;
WATTS, O ;
SHROOT, B ;
JAMOULLE, JC .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1989, 49 (01) :63-68
[4]  
RITSCHEL WA, 1989, METHOD FIND EXP CLIN, V11, P165