THE OPERATION OF MICROCHANNEL PLATES AT HIGH COUNT RATES

被引:60
作者
FRASER, GW
PAIN, MT
LEES, JE
PEARSON, JF
机构
[1] X-ray Astronomy Group, Department of Physics and Astronomy, University of Leicester, Leicester, LE1 7RH England, University Road
关键词
D O I
10.1016/0168-9002(91)90329-O
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The measured count rate characteristics of a large number of microchannel plate electron multipliers are compared with the predictions of a universal, paralysable-counter model. Individual plate resistances for the experimental study lie in the range 27-2450 M-OMEGA. The gain behaviour of single channel plates is shown to differ from that of multi-stage MCP detectors. The measured dependence of multi-stage multiplier recovery time on illuminated area is interpreted in terms of inter-channel coupling and of changes in conduction current during plate operation. Our findings have significance for the calibration of particle and photon spectrometers and for the development of future channel plate detectors with extended dynamic range.
引用
收藏
页码:247 / 260
页数:14
相关论文
共 34 条
[1]  
AUDIER M, 1975, PHILIPS RES REP, V30, P226
[2]   NEW PHOTOMULTIPLIER TUBE UTILIZING CHANNEL PLATE ELECTRON MULTIPLIERS AS GAIN PRODUCING ELEMENTS [J].
BATEMAN, JE ;
APSIMON, RJ ;
BARLOW, FE .
NUCLEAR INSTRUMENTS & METHODS, 1976, 137 (01) :61-70
[3]  
BLATOV FI, 1990, INSTRUM EXP TECH+, V33, P647
[4]  
CHIAPPETTI L, 1974, ESA EXOSAT EXPRESS
[5]  
Cho D.J., 1988, P SPIE, V976, P172, DOI [10.1117/12.948543, DOI 10.1117/12.948543]
[6]  
CHO DJ, 1989, THESIS U ROCHESTER
[7]  
DEBRUIN DP, 1920, REV SCI INSTRUM, V53, P1020
[8]   AN OPERATIONAL MODEL FOR MICROCHANNEL PLATE DEVICES [J].
EBERHARDT, EH .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1981, 28 (01) :712-717
[9]  
EBERHARDT EH, 1980, ITT127 EL OPT PROD D
[10]  
FELLER WB, 1990, P SOC PHOTO-OPT INS, V1243, P100