NOVEL FIA AMPEROMETRIC BIOSENSOR SYSTEM FOR THE DETERMINATION OF GLUTAMINE IN CELL-CULTURE SYSTEMS

被引:15
作者
MALE, KB [1 ]
LUONG, JHT [1 ]
TOM, R [1 ]
MERCILLE, S [1 ]
机构
[1] NATL RES COUNCIL CANADA,BIOTECHNOL RES INST,MONTREAL H4P 2R2,QUEBEC,CANADA
关键词
BIOSENSOR SYSTEM; FLOW INJECTION ANALYSIS; GLUTAMINE; GLUTAMATE; GLUTAMINASE; GLUTAMATE OXIDASE; ANION EXCHANGE RESINS; MAMMALIAN CELLS; INSECT CELLS;
D O I
10.1016/0141-0229(93)90112-F
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
A flow injection analysis (FIA) biosensor system has been developed for determining glutamine in insect cell and in murine hybridoma cell cultures. Glutamate oxidase and glutaminase, respectively, were covalently immobilized onto porous aminopropyl glass beads to form an immobilized enzyme column. The hydrogen peroxide produced by the enzyme reactions was detected by an amperometric electrode (platinum vs. silver/silver chloride) poised at + 0.7 V. Among several anion exchange resins tested, endogenous glutamate was completely adsorbed onto acetate anion exchange resins. Such an ion exchanger also effectively adsorbed aspartic acid, uric acid, ascorbic acid, and acetaminophen. The FIA biosensor was linear up to 1 mM glutamine, with a lower detection limit of 10 muM, and possessed good reproducibility (relative error of +/- 1.2%). Each analysis could be performed in 3.5 min including sampling and washing with a corresponding throughput of 17 h-1. The anion exchange column could be operated continuously for 12 h for 200 analyses with diluted cell culture. The immobilized enzyme column was stable for at least 500 repeated analyses without significant loss of activity. When the biosensor system was applied to glutamine measurement in insect cell and mammalian cell cultures, the results obtained compared well with those HPLC.
引用
收藏
页码:26 / 32
页数:7
相关论文
共 13 条
[1]  
ADAMSON SR, 1987, COMMERCIAL PRODUCTIO, P17
[2]   MONITORING GLUTAMINE IN MAMMALIAN-CELL CULTURES USING AN AMPEROMETRIC BIOSENSOR [J].
CATTANEO, MV ;
LUONG, JHT ;
MERCILLE, S .
BIOSENSORS & BIOELECTRONICS, 1992, 7 (05) :329-334
[3]  
COCHRAN DG, 1975, INSECT BIOCH FUNCTIO, P179
[4]   THE EFFECT OF PH ON THE TOXICITY OF AMMONIA TO A MURINE HYBRIDOMA [J].
DOYLE, C ;
BUTLER, M .
JOURNAL OF BIOTECHNOLOGY, 1990, 15 (1-2) :91-100
[5]  
Ham R G, 1979, Methods Enzymol, V58, P44
[6]   ELECTROCHEMICAL PLATINIZATION OF RETICULATED VITREOUS CARBON ELECTRODES TO INCREASE BIOSENSOR RESPONSE [J].
HEIDER, GH ;
SASSO, SV ;
HUANG, KM ;
YACYNYCH, AM ;
WIECK, HJ .
ANALYTICAL CHEMISTRY, 1990, 62 (11) :1106-1110
[7]   PURIFICATION AND PROPERTIES OF A NEW ENZYME, L-GLUTAMATE OXIDASE, FROM STREPTOMYCES SP X-119-6 GROWN ON WHEAT BRAN [J].
KUSAKABE, H ;
MIDORIKAWA, Y ;
FUJISHIMA, T ;
KUNINAKA, A ;
YOSHINO, H .
AGRICULTURAL AND BIOLOGICAL CHEMISTRY, 1983, 47 (06) :1323-1328
[8]  
LEHNINGER AL, 1982, PRINCIPLES BIOCH, P107
[9]   DETERMINATION OF ASPARTAME IN DIETARY FOOD-PRODUCTS BY A FIA BIOSENSOR [J].
MALE, KB ;
LUONG, JHT ;
MULCHANDANI, A .
BIOSENSORS & BIOELECTRONICS, 1991, 6 (02) :117-123
[10]   THEORY AND APPLICATION OF DIFFUSION-LIMITED AMPEROMETRIC ENZYME ELECTRODE DETECTION IN FLOW-INJECTION ANALYSIS OF GLUCOSE [J].
OLSSON, B ;
LUNDBACK, H ;
JOHANSSON, G ;
SCHELLER, F ;
NENTWIG, J .
ANALYTICAL CHEMISTRY, 1986, 58 (06) :1046-1052