EVALUATION OF A SUBCUTANEOUS GLUCOSE SENSOR OUT TO 3 MONTHS IN A DOG-MODEL

被引:72
作者
GILLIGAN, BJ
SHULTS, MC
RHODES, RK
UPDIKE, SJ
机构
[1] UNIV WISCONSIN, CTR HLTH SCI, DEPT MED H4530, BIOMED ENGN LAB, MADISON, WI 53792 USA
[2] MARKWELL MED INST, RACINE, WI USA
关键词
D O I
10.2337/diacare.17.8.882
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVE - To advance the feasibility of an implantable long-term glucose sensor with bioprotective sensor membranes and test protocols using a somatostatin analog (octreotide). RESEARCH DESIGN AND METHODS - Implantable sensors were constructed with one of eight bioprotective membranes and screened in vitro for stable response to glucose. Sensors were implanted subcutaneously into nondiabetic mongrel dogs and monitored at 4-min intervals via radiotelemetry. When implanted sensor responses showed evidence of tracking blood glucose after glucagon challenge (8-21 days postimplant), a glucose infusion protocol was used to assess performance. Sensor data were collected every 4 s after octreotide inhibition of endogenous insulin release. Reference plasma glucose samples were taken every 4-10 min. RESULTS - Preimplant in vitro testing of sensors verified linearity to 33.3 mM glucose and response times to 90% of equilibrium in 2-7 min. Ten implanted sensors tracked glucose for 20-114 days, during which 25 separate glucose infusion studies were conducted. The resulting regression data yielded a mean slope of 0.99 +/- 0.06, an intercept of 0.24 +/- 0.53 mM glucose, and a correlation coefficient 0.98 +/- 0.01. Long-term sensor stability was not judged adequate for clinical application, although two sensors tracked within +/- 15% for 33 and 42 days. In vivo oxygen delivery was shown to affect sensor performance. On explant, two of eight tested bioprotective membranes were found to be biostable and to fully protect the sensor's enzyme membrane. The foreign body capsule was adequately vascularized adjacent to the sensor up to 91 days postimplant. Sensor units eventually failed because of electronic problems (package leakage) or because of biodegradation or biofouling of test bioprotective membranes. CONCLUSION - Further development of this type of sensor sensor may provide diabetic patients with a better means of monitoring blood glucose.
引用
收藏
页码:882 / 887
页数:6
相关论文
共 18 条
  • [1] IMPORTANCE OF GLUCOSE PER SE TO INTRAVENOUS GLUCOSE-TOLERANCE - COMPARISON OF THE MINIMAL-MODEL PREDICTION WITH DIRECT MEASUREMENTS
    ADER, M
    PACINI, G
    YANG, YJ
    BERGMAN, RN
    [J]. DIABETES, 1985, 34 (11) : 1092 - 1103
  • [2] APPLICATION OF CHRONIC INTRAVASCULAR BLOOD-GLUCOSE SENSOR IN DOGS
    ARMOUR, JC
    LUCISANO, JY
    MCKEAN, BD
    GOUGH, DA
    [J]. DIABETES, 1990, 39 (12) : 1519 - 1526
  • [3] BRAUKER J, 1992, 4TH P WORLD BIOM C B
  • [4] GLYCATION PRODUCTS AND THE PATHOGENESIS OF DIABETIC COMPLICATIONS
    BROWNLEE, M
    [J]. DIABETES CARE, 1992, 15 (12) : 1835 - 1843
  • [5] CLARK LC, 1988, T AM SOC ART INT ORG, V34, P259
  • [6] DCCT Res Grp, 1991, AM J MED, V90, P450, DOI 10.1016/0002-9343(91)80085-Z
  • [7] FERRANNINI E, 1986, METHODS DIABETIC RES, V2, P111
  • [8] GREISLER HP, 1992, NEW BIOL SYNTHETIC V, P77
  • [9] Phillips R E, 1988, J Biomater Appl, V3, P207, DOI 10.1177/088532828800300204
  • [10] PREDICTION OF POCKET-PORTABLE AND IMPLANTABLE GLUCOSE ENZYME ELECTRODE PERFORMANCE FROM COMBINED SPECIES PERMEABILITY AND DIGITAL-SIMULATION ANALYSIS
    RHODES, RK
    SHULTS, MC
    UPDIKE, SJ
    [J]. ANALYTICAL CHEMISTRY, 1994, 66 (09) : 1520 - 1529