SPECTRAL CHARACTERISTICS OF CHEST-WALL BREATH SOUNDS IN NORMAL SUBJECTS

被引:88
作者
GAVRIELY, N
NISSAN, M
RUBIN, AHE
CUGELL, DW
机构
[1] TECHNION ISRAEL INST TECHNOL,RAPPAPORT FAMILY INST RES MED SCI,BRUCE RAPPAPORT FAC MED,IL-31096 HAIFA,ISRAEL
[2] TECHNION ISRAEL INST TECHNOL,INST WORK SAFETY & HUMAN ENGN,HAIFA,ISRAEL
[3] RAMBAM MED CTR,DIV PULM,HAIFA,ISRAEL
[4] NORTHWESTERN UNIV,SCH MED,DEPT ANESTHESIA,CHICAGO,IL 60611
[5] NORTHWESTERN UNIV,SCH MED,PULM SECT,CHICAGO,IL 60611
[6] NORTHWESTERN UNIV,NW MEM HOSP,CHICAGO,IL 60611
关键词
PHONOPNEUMOGRAPHY; LUNG SOUNDS; NORMAL STANDARDS; PULMONARY ACOUSTICS; SMOKERS; SEX;
D O I
10.1136/thx.50.12.1292
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background - This study was carried out to establish a reliable bank of information on the spectral characteristics of chest wall breath sounds from healthy men and women, both non-smokers and smokers. Methods - Chest wall breath sounds from 272 men and 81 women were measured using contact acoustic sensors, amplifiers, and fast Fourier transform (FFT) based spectral analysis software. Inspiratory and expiratory sounds were picked up at three standard locations on the chest wall during breathing at flows of 1-2 Ys and analysed breath by breath in real time. Results - The amplitude spectrum of normal chest wall breath sounds has two linear parts in the log-log plane - low and high frequency segments - that are best characterised by their corresponding regression lines. Four parameters are needed and are sufficient for complete quantitative representation of each of the spectra: the slopes of the two regression lines plus the amplitude and frequency coordinates of their intersection. The range of slopes of the high frequency lines was -12.7 to -15.2 dB/oct during inspiration and -13.4 to -20.3 dB/oct during expiration. The frequency at which this Line crossed the zero dB level - that is, the amplitude resolution threshold of the system - was designated as the maximal Technology, frequency (Fmax) which varied from 736 to 999 Hz during inspiration and from 426 to 796 Hz during expiration with higher values in women than in men. The mean (SD) regression coefficient of the high frequency line was 0.89 (0.05). Conclusions - These data define the boundaries of normal chest wall breath sounds and may be used as reference for comparison with abnormal sounds.
引用
收藏
页码:1292 / 1300
页数:9
相关论文
共 15 条
[1]   VARIATION OF BREATH SOUND AND AIRWAY CALIBER INDUCED BY HISTAMINE CHALLENGE [J].
ANDERSON, K ;
AITKEN, S ;
CARTER, R ;
MACLEOD, JES ;
MORAN, F .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1990, 141 (05) :1147-1150
[2]   ANALYSIS AND AUTOMATIC CLASSIFICATION OF BREATH SOUNDS [J].
COHEN, A ;
LANDSBERG, D .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1984, 31 (09) :585-590
[3]   RESPIRATORY HEALTH SCREENING USING PULMONARY-FUNCTION TESTS AND LUNG SOUND ANALYSIS [J].
GAVRIELY, N ;
NISSAN, M ;
CUGELL, DW ;
RUBIN, AHE .
EUROPEAN RESPIRATORY JOURNAL, 1994, 7 (01) :35-42
[4]   SPECTRAL CHARACTERISTICS OF NORMAL BREATH SOUNDS [J].
GAVRIELY, N ;
PALTI, Y ;
ALROY, G .
JOURNAL OF APPLIED PHYSIOLOGY, 1981, 50 (02) :307-314
[5]   PARAMETRIC REPRESENTATION OF NORMAL BREATH SOUNDS [J].
GAVRIELY, N ;
HERZBERG, M .
JOURNAL OF APPLIED PHYSIOLOGY, 1992, 73 (05) :1776-1784
[6]  
Gavriely N, 1995, BREATH SOUNDS METHOD, DOI 10.1201/9780429260544
[7]  
GAVRIELY N, 1992, 17TH INT C LUNG SOUN
[8]  
LAIRD CW, 1974, LANCET, V1, P808
[9]  
MACLEOD JG, 1976, CLIN EXAMINATION
[10]   VISUAL LUNG-SOUND CHARACTERIZATION BY TIME-EXPANDED WAVE-FORM ANALYSIS [J].
MURPHY, RLH ;
HOLFORD, SK ;
KNOWLER, WC .
NEW ENGLAND JOURNAL OF MEDICINE, 1977, 296 (17) :968-971