WHITE CELL DEPLETION OF SINGLE-DONOR PLATELET PREPARATIONS BY A NEW ADSORPTION FILTER

被引:7
作者
BOCK, M
HEIM, MU
WEINDLER, R
BILAS, A
GREITHER, L
SALAT, C
MEMPEL, W
机构
[1] Transfusion Center, Klinikum Grosshadern, University of Munich, Munich
关键词
D O I
10.1046/j.1537-2995.1991.31491213298.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A newly developed adsorption filter has been evaluated for the removal of white cells (WBCs) from platelet concentrates. It consists of surface-modified polyester fibers and can be used at the patient's bedside. WBC mean removal rate was 98.8 +/- 0.9 percent, and platelet recovery was 99.0 +/- 0.7 percent. The aggregation curves (ADP, collagen) and hypotonic shock response of platelets were not influenced by the filtration, and the plasma electrolyte and lactate dehydrogenase concentrations were unaffected. Transmission and scanning electron micrographs showed no difference in platelet ultrastructure before and after filtration. Determination of posttransfusion platelet recovery did not reveal any significant difference from that of nonfiltered platelets. The tested filter seems to be a suitable device for WBC depletion of single-donor platelet concentrates at the bedside.
引用
收藏
页码:333 / 334
页数:2
相关论文
共 13 条
[11]   PREPARATION OF LEUKOCYTE-FREE PLATELETS FOR TRANSFUSION BY FILTRATION THROUGH COTTON WOOL [J].
SIRCHIA, G ;
PARRAVICINI, A ;
REBULLA, P ;
BERTOLINI, F ;
MORELATI, F ;
MARCONI, M .
VOX SANGUINIS, 1983, 44 (02) :115-120
[12]   USE OF POLYESTER FILTERS FOR THE PREPARATION OF LEUKOCYTE-POOR PLATELET CONCENTRATES [J].
SNYDER, EL ;
DEPALMA, L ;
NAPYCHANK, P .
VOX SANGUINIS, 1988, 54 (01) :21-23
[13]  
VANPROOIJEN HC, 1987, BLOOD, V70, P243