THE EFFECT OF ECHO SUPPRESSION ON THE MEAN VELOCITY ESTIMATION RANGE OF THE RF CROSS-CORRELATION MODEL ESTIMATOR

被引:25
作者
BRANDS, PJ [1 ]
HOEKS, APG [1 ]
RENEMAN, RS [1 ]
机构
[1] UNIV LIMBURG,CARIM,DEPT PHYSIOL,6200 MD MAASTRICHT,NETHERLANDS
关键词
CLUTTER SUPPRESSION; PULSED DOPPLER; RF CROSS-CORRELATION; RF-DOMAIN SIGNAL PROCESSING; ULTRASOUND; VESSEL WALL FILTER;
D O I
10.1016/0301-5629(95)00028-P
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Ultrasound systems are widely used to visualize in real-time internal structures and blood flow velocity distributions. The latter are estimated from the received ultrasound radio frequency (RIF) signals using a mean frequency estimator, capable of estimating the temporal and spatial mean frequency within a given depth and time window (estimation window), Since ultrasound RF-signals are composed of three major signal components (reflections, scattering and noise) it is necessary to suppress the power of the reflections or reverberations (stationary echo removal) to estimate the temporal mean frequency of the signal component induced by scattering originating from red blood cells, However, the stationary echo removal filter in front of the mean frequency estimator will restrict the temporal mean frequency estimation range because the power of the signal component induced by the slowly moving blood is suppressed as well. This article describes a stationary echo removal filter, acting in the RF-domain and adaptive to the temporal mean frequency of the signal component induced by stationary or slowly moving reflections and reverberations, Furthermore, a comparison in performance of this adaptive stationary echo removal filter and several static stationary echo removal filters is presented. For direct comparison of these filter algorithms as far as the effect on the temporal mean frequency estimation range is concerned, the same signal conditions in combination with the same RF-domain mean frequency estimator were used in this comparison. It could be demonstrated that the adaptive echo removal filter, in combination with the RF-domain mean frequency estimator used, exhibits the best performance.
引用
收藏
页码:945 / 959
页数:15
相关论文
共 20 条
[1]   DISCRETE-TIME ESTIMATION OF THE MEAN DOPPLER FREQUENCY IN ULTRASONIC BLOOD VELOCITY-MEASUREMENTS [J].
ANGELSEN, BAJ ;
KRISTOFFERSEN, K .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (04) :207-214
[2]   TIME-DELAY ESTIMATION BY GENERALIZED CROSS-CORRELATION METHODS [J].
AZARIA, M ;
HERTZ, D .
IEEE TRANSACTIONS ON ACOUSTICS SPEECH AND SIGNAL PROCESSING, 1984, 32 (02) :280-285
[3]  
Brandestini MA, 1979, ECHOCARDIOLOGY, P441
[4]   A NONINVASIVE METHOD TO ESTIMATE WALL SHEAR RATE USING ULTRASOUND [J].
BRANDS, PJ ;
HOEKS, APG ;
HOFSTRA, L ;
RENEMAN, RS .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (02) :171-185
[5]   A COMPARISON METHOD FOR MEAN FREQUENCY ESTIMATORS FOR DOPPLER ULTRASOUND [J].
BRANDS, PJ ;
HOEKS, APG .
ULTRASONIC IMAGING, 1992, 14 (04) :367-386
[6]   DETERMINATION OF TISSUE MOTION VELOCITY BY CORRELATION INTERPOLATION OF PULSED ULTRASONIC ECHO SIGNALS [J].
DEJONG, PGM ;
ARTS, T ;
HOEKS, APG ;
RENEMAN, RS .
ULTRASONIC IMAGING, 1990, 12 (02) :84-98
[7]   A NEW WIDE-BAND SPREAD TARGET MAXIMUM-LIKELIHOOD ESTIMATOR FOR BLOOD VELOCITY ESTIMATION .1. THEORY [J].
FERRARA, KW ;
ALGAZI, VR .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1991, 38 (01) :1-16
[8]   THE EFFECT OF FREQUENCY-DEPENDENT SCATTERING AND ATTENUATION ON THE ESTIMATION OF BLOOD VELOCITY USING ULTRASOUND [J].
FERRARA, KW ;
ALGAZI, VR ;
LIU, JX .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1992, 39 (06) :754-767
[9]   A MULTIGATE PULSED DOPPLER SYSTEM WITH SERIAL DATA-PROCESSING [J].
HOEKS, APG ;
RENEMAN, RS ;
PERONNEAU, PA .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1981, 28 (04) :242-247
[10]   COMPARISON OF THE PERFORMANCE OF THE RF CROSS-CORRELATION AND DOPPLER AUTOCORRELATION TECHNIQUE TO ESTIMATE THE MEAN VELOCITY OF SIMULATED ULTRASOUND SIGNALS [J].
HOEKS, APG ;
ARTS, TGJ ;
BRANDS, PJ ;
RENEMAN, RS .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1993, 19 (09) :727-740