SOME PROBLEMS IN PRECISION THERMOLUMINESCENCE DOSIMETRY

被引:10
作者
ZARAND, P
机构
[1] National Institute of Roentgen and Radiation Physics, Budapest
关键词
D O I
10.1088/0031-9155/24/6/019
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In low dose measurements the statistical nature of the light emission and light conversion as well as that of the photomultiplication sets an absolute limit for the accuracy of a particular TL reading. The standard deviation achieved with LiF pellets is greater by a factor of about 2.5 than the optimal value determined by using a light source with variable intensity. The statistical fluctuation of the dark current determines the lower detection limit of the particular read-out instrument. Above a certain dose, in the present experiments about 0.1 Gy, the effect of statistical fluctuations observed in the low dose range is negligible and the precision of the measurements is determined solely by the reproducibility or the statistical uncertainty for a repeated measurement of the same exposure.
引用
收藏
页码:1284 / 1290
页数:7
相关论文
共 27 条
[1]  
BOTTERJENSEN L, 1971, ADV PHYSICAL BIOL RA, P113
[2]   USEFULNESS OF INTERCOMPARISON STUDIES FOR IMPROVEMENT OF X-RAY DOSIMETRY [J].
BROERSE, JJ ;
PUITE, KJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1974, 19 (05) :732-734
[3]  
BRUNSKILL RT, 1971, ADV PHYSICAL BIOL RA, P67
[4]  
CAMERON J. R., 1964, HLTH PHYS, V10, P25, DOI 10.1097/00004032-196401000-00003
[5]   THERMOLUMINESCENCE OF LIF - DEPENDENCE OF THERMAL HISTORY [J].
CARLSSON, CA .
PHYSICS IN MEDICINE AND BIOLOGY, 1969, 14 (01) :107-&
[6]   IAEA-WHO CO-60 TELETHERAPY DOSIMETRY SERVICE USING MAILED LIF DOSIMETERS - SURVEY OF RESULTS OBTAINED DURING 1970-75 [J].
EISENLOHR, HH ;
JAYARAMAN, S .
PHYSICS IN MEDICINE AND BIOLOGY, 1977, 22 (01) :18-28
[7]  
LARIVIERE PD, 1976, HEALTH PHYS, V30, P409
[8]   SENSITIZATION IN LIF-TEFLON DOSEMETERS [J].
LINSLEY, GS ;
MASON, EW .
PHYSICS IN MEDICINE AND BIOLOGY, 1971, 16 (04) :695-&
[9]  
MAKRA Z, 1977, OPUT PRIMENENIA INDI, P36
[10]   THERMOLUMINESCENCE OF LIF - A STATISTICAL ANALYSIS OF INFLUENCE OF PRE-ANNEALING ON PRECISION OF MEASUREMENT [J].
MARTENSS.BK .
PHYSICS IN MEDICINE AND BIOLOGY, 1969, 14 (01) :119-&