DETECTION AND IMAGING OF ELECTRIC-CONDUCTIVITY AND PERMITTIVITY AT LOW-FREQUENCY

被引:45
作者
FUKS, LF
CHENEY, M
ISAACSON, D
GISSER, DG
NEWELL, JC
机构
[1] RENSSELAER POLYTECH INST, DEPT MATH SCI, TROY, NY 12180 USA
[2] RENSSELAER POLYTECH INST, DEPT BIOMED ENGN, TROY, NY 12180 USA
[3] RENSSELAER POLYTECH INST, DEPT COMP ENGN, TROY, NY 12180 USA
[4] RENSSELAER POLYTECH INST, DEPT SYST ENGN, TROY, NY 12180 USA
[5] UNION UNIV, PHYSIOL & SURG, ALBANY, NY 12208 USA
关键词
D O I
10.1109/10.99074
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper is concerned with low frequency electrical impedance imaging, which is the process of constructing images of the electrical impedance of a body's interior based upon measurements of voltage and current made at the body's surface. The electrical impedance accounts for both resistivity and permittivity. This paper shows how permittivity can be exploited to improve the performance of an electrical impedance imaging system. We show that explicit use of the independent information in the data due to the permittivity will ehance a system's ability to distinguish objects in the interior of a body. In addition, we report the results of experiments performed using the Rensselaer ACT 2 system on a saline bath containing various objects. These objects include both living tissue and metal conductors with oxide layers. We demonstrate the system's ability to distinguish these objects, and we exhibit gray scale images of both their resistivity and permittivity distributions.
引用
收藏
页码:1106 / 1110
页数:5
相关论文
共 23 条
  • [1] Barber D C, 1987, Clin Phys Physiol Meas, V8 Suppl A, P47, DOI 10.1088/0143-0815/8/4A/006
  • [2] Brown B H, 1987, Clin Phys Physiol Meas, V8 Suppl A, P91, DOI 10.1088/0143-0815/8/4A/012
  • [3] APPLIED POTENTIAL TOMOGRAPHY - POSSIBLE CLINICAL-APPLICATIONS
    BROWN, BH
    BARBER, DC
    SEAGAR, AD
    [J]. CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1985, 6 (02): : 109 - 121
  • [4] FOSTER KR, 1989, CRIT REV BIOMED ENG, V17, P25
  • [5] REACTIVE EFFECTS IN IMPEDANCE IMAGING
    FUKS, LF
    ISAACSON, D
    GISSER, DG
    NEWELL, JC
    [J]. PROCEEDINGS OF THE ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, PTS 1-4, 1988, : 281 - 282
  • [6] FUKS LF, 1989, THESIS RENSSELAER PO
  • [7] Gisser D G, 1987, Clin Phys Physiol Meas, V8 Suppl A, P39, DOI 10.1088/0143-0815/8/4A/005
  • [8] GOBLE JC, 1988, P IEEE ENG MED BIOL
  • [9] THE IMPORTANCE OF PHASE MEASUREMENT IN ELECTRICAL-IMPEDANCE TOMOGRAPHY
    GRIFFITHS, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (11) : 1435 - 1444
  • [10] GRIFFITHS H, 1989, P IEEE ENG MED BIOL, P476