THE APPLICATION OF EXPERIMENTAL-DESIGN TO THE ROBUSTNESS TESTING OF A METHOD FOR THE DETERMINATION OF DRUG-RELATED IMPURITIES BY CAPILLARY ELECTROPHORESIS

被引:42
作者
ALTRIA, KD
FILBEY, SD
机构
[1] Pharmaceutical Analysis, Glaxo Research and Development, Ware, SG12 0DP, Herts, Park Road
关键词
CAPILLARY ELECTROPHORESIS; EXPERIMENTAL DESIGN; ROBUSTNESS TESTING; PHARMACEUTICAL ANALYSIS;
D O I
10.1007/BF02274518
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Robustness testing of capillary electrophoresis methods is an important part of method validation. Appropriate use of experimental designs can be employed in this robustness testing. In this study experimental designs were used in the assessment of a capillary electrophoresis method used to determine drug related impurities. Initially a fractional factorial screening design was used to identify the critical parameters which were found to be pH and voltage. A central composite design was then performed to evaluate the response surfaces for pH and voltage which showed operation at the optimum pH and voltage values.
引用
收藏
页码:306 / 310
页数:5
相关论文
共 12 条
  • [1] INTER-COMPANY CROSS-VALIDATION EXERCISE ON CAPILLARY ELECTROPHORESIS .1. CHIRAL ANALYSIS OF CLENBUTEROL
    ALTRIA, KD
    HARDEN, RC
    HART, M
    HEVIZI, J
    HAILEY, PA
    MAKWANA, JV
    PORTSMOUTH, MJ
    [J]. JOURNAL OF CHROMATOGRAPHY, 1993, 641 (01): : 147 - 153
  • [2] ESSENTIAL PEAK AREA NORMALIZATION FOR QUANTITATIVE IMPURITY CONTENT DETERMINATION BY CAPILLARY ELECTROPHORESIS
    ALTRIA, KD
    [J]. CHROMATOGRAPHIA, 1993, 35 (3-4) : 177 - 182
  • [4] EFFECT OF ELECTROLYTE AND SAMPLE CONCENTRATION ON THE RELATIONSHIP BETWEEN SENSITIVITY AND RESOLUTION IN CAPILLARY ZONE ELECTROPHORESIS USING CONDUCTIVITY DETECTION
    HUANG, XH
    GORDON, MJ
    ZARE, RN
    [J]. JOURNAL OF CHROMATOGRAPHY, 1989, 480 : 285 - 288
  • [5] DETERMINATION OF CYTOSINE-BETA-D-ARABINOSIDE IN PLASMA USING CAPILLARY ELECTROPHORESIS
    LLOYD, DK
    CYPESS, AM
    WAINER, IW
    [J]. JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1991, 568 (01): : 117 - 124
  • [6] RUGGEDNESS TESTING IN ANALYTICAL-CHEMISTRY
    MULHOLLAND, M
    [J]. TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1988, 7 (10) : 383 - 389
  • [7] CAPILLARY ELECTROPHORESIS IN CHEMICAL PHARMACEUTICAL QUALITY-CONTROL
    PLUYM, A
    VANAEL, W
    DESMET, M
    [J]. TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1992, 11 (01) : 27 - 32
  • [8] EVALUATION OF A MIXED MICELLAR ELECTROKINETIC CAPILLARY ELECTROPHORESIS METHOD FOR VALIDATED PHARMACEUTICAL QUALITY-CONTROL
    THOMAS, BR
    GHODBANE, S
    [J]. JOURNAL OF LIQUID CHROMATOGRAPHY, 1993, 16 (9-10): : 1983 - 2006
  • [9] RESOLUTION OPTIMIZATION IN MICELLAR ELECTROKINETIC CHROMATOGRAPHY - USE OF PLACKETT-BURMAN STATISTICAL DESIGN FOR THE ANALYSIS OF TESTOSTERONE ESTERS
    VINDEVOGEL, J
    SANDRA, P
    [J]. ANALYTICAL CHEMISTRY, 1991, 63 (15) : 1530 - 1536
  • [10] RATIONAL DESIGN OF POTENT SIALIDASE-BASED INHIBITORS OF INFLUENZA-VIRUS REPLICATION
    VONITZSTEIN, M
    WU, WY
    KOK, GB
    PEGG, MS
    DYASON, JC
    JIN, B
    PHAN, TV
    SMYTHE, ML
    WHITE, HF
    OLIVER, SW
    COLMAN, PM
    VARGHESE, JN
    RYAN, DM
    WOODS, JM
    BETHELL, RC
    HOTHAM, VJ
    CAMERON, JM
    PENN, CR
    [J]. NATURE, 1993, 363 (6428) : 418 - 423