ELECTROCARDIOGRAPHIC ARTIFACT WITH A TRANSCUTANEOUS ELECTRICAL NERVE-STIMULATION UNIT

被引:13
作者
HAUPTMAN, PJ
RAZA, M
机构
[1] Department of Cardiology, Mount Sinai Hospital, New York, NY
关键词
ELECTROCARDIOGRAM; ARTIFACT;
D O I
10.1016/0167-5273(92)90093-I
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We describe an unusual electrocardiographic artifact in a patient with a transcutaneous electrical nerve stimulation unit. The artifact, consisting of low amplitude high frequency deflections, can be misinterpreted as a runaway pacemaker.
引用
收藏
页码:110 / 112
页数:3
相关论文
共 5 条
[1]  
CHUGN E, 1980, ELECTROCARDIOGRAPHY, P641
[2]  
KIMBERLEY APS, 1987, ANAESTH INTENS CARE, V15, P358
[3]   ECG ARTIFACTS DURING INTRAOPERATIVE EVOKED-POTENTIAL MONITORING [J].
LEGATT, AD ;
FROST, EAM .
ANESTHESIOLOGY, 1989, 70 (03) :559-559
[4]   ELECTROCARDIOGRAPHIC ARTIFACT DUE TO MALFUNCTION OF AN ELECTRIC WATCH [J].
LESCH, M ;
GREENE, HL .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1974, 228 (01) :26-26
[5]  
YANG SS, 1983, GUIDE CARDIAC PACEME, P71