COMPARISON OF PROCEDURES FOR OBTAINING THRESHOLDS AND MAXIMUM ACCEPTABLE LOUDNESS LEVELS WITH THE NUCLEUS COCHLEAR IMPLANT SYSTEM

被引:22
作者
SKINNER, MW [1 ]
HOLDEN, LK [1 ]
HOLDEN, TA [1 ]
DEMOREST, ME [1 ]
机构
[1] UNIV MARYLAND,BALTIMORE,MD 21201
来源
JOURNAL OF SPEECH AND HEARING RESEARCH | 1995年 / 38卷 / 03期
关键词
COCHLEAR IMPLANT; THRESHOLD; MAXIMUM ACCEPTABLE LOUDNESS; BEHAVIORAL PROCEDURES;
D O I
10.1044/jshr.3803.677
中图分类号
H [语言、文字];
学科分类号
05 ;
摘要
Two stimulus paradigms and two presentation methods were combined to form three procedures (keyboard, knob, and ascending loudness judgments with know [ALJK]) to obtain detection thresholds and maximum acceptable loudness levels (MALs) from 11 adults with the Nucleus cochlear implant. Thresholds at which subjects correctly counted the number of stimulus bursts also were obtained. Keyboard detection thresholds were higher (Scheffe, p = .01) than knob and ALJK detection thresholds. Counted thresholds were obtained most efficiently by using keyboard detection thresholds as the initial level for testing. Keyboard MALs were highest, knob MALs intermediate, and ALJK MALs lowest (Scheffe, p = .001). MALs were obtained most efficiently with the ALJK procedure. Implications of these results for clinical practice are discussed.
引用
收藏
页码:677 / 689
页数:13
相关论文
共 14 条
[1]  
BONDING P, 1979, AUDIOLOGY, V18, P197
[2]   PREFERRED METHOD FOR CLINICAL DETERMINATION OF PURE-TONE THRESHOLDS [J].
CARHART, R ;
JERGER, JF .
JOURNAL OF SPEECH AND HEARING DISORDERS, 1959, 24 (04) :330-345
[3]  
CLARK GM, 1987, ADV OTORHINOLARYNGOL, V38, P1
[4]  
*COCHL CORP, 1989, AUD HDB
[5]  
DILLIER N, 1993, PROGR BRAIN RES, V97, P303
[6]  
KHALIL E, 1989, ASHA, V31, P126
[7]  
KWIECINSKI BG, 1983, STUDY THRESHOLD LOUD
[8]  
MATSUMOTO J, 1983, UNPUB STUDY THRESHOL
[9]  
MCNEMAR Q, 1962, PSYCHOL STATISTICS
[10]   EFFECTS OF PHASE DURATION ON DETECTION OF ELECTRICAL-STIMULATION OF THE HUMAN COCHLEA [J].
MOON, AK ;
ZWOLAN, TA ;
PFINGST, BE .
HEARING RESEARCH, 1993, 67 (1-2) :166-178