DETERMINATION OF D-GLUCOSE IN UNDILUTED WHOLE-BLOOD USING CHEMICALLY MODIFIED ELECTRODES AND SEGMENTED SAMPLE INJECTION IN A FLOW SYSTEM

被引:7
作者
BUCHRASMUSSEN, T [1 ]
机构
[1] RADIOMETER MED A-S,DK-2400 COPENHAGEN,DENMARK
关键词
Blood; Enzyme reactor; Glucose;
D O I
10.1016/S0003-2670(00)83944-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow system for the determination of d-glucose in undiluted whole blood, in which segmented sample injection and on-line dialysis are used to decrease the red cell volume fraction (haematrocrit) dependence, is described. Glucose is degraded enzymatically by means of immobilized glucose dehydrogenase. The nicotinamide coenzyme (NAD+) that is present in the solution is reduced in the enzymatic reaction and is reoxidized amperometically at 0 mV vs. Ag/AgCl on a graphite electrode, modified with phenoxazinium ion. The potential use of the system for clinical analysis is evaluated. © 1990.