INCORPORATING THE GAS ANALYZER RESPONSE-TIME IN GAS-EXCHANGE COMPUTATIONS

被引:33
作者
MITCHELL, RR
机构
关键词
D O I
10.1152/jappl.1979.47.5.1118
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
A simple method for including the gas analyzer response time in the breath-by-breath computation of gas exchange rates is described. The method uses a difference equation form of a model for the gas analyzer in the computation of oxygen uptake and carbon dioxide production and avoids a numerical differentiation required to correct the gas fraction wave forms. The effect of not accounting for analyzer response time is shown to be a 20% underestimation in gas exchange rate. The present method accurately measures gas exchange rate, is relatively insensitive to measurement errors in the analyzer time constant, and does not significantly increase the computation time.
引用
收藏
页码:1118 / 1122
页数:5
相关论文
共 12 条
[1]   ON-LINE COMPUTER ANALYSIS AND BREATH-BY-BREATH GRAPHICAL DISPLAY OF EXERCISE FUNCTION TESTS [J].
BEAVER, WL ;
WASSERMAN, K ;
WHIPP, BJ .
JOURNAL OF APPLIED PHYSIOLOGY, 1973, 34 (01) :128-132
[2]   ASPECTS OF ONLINE DIGITAL INTEGRATION OF PULMONARY GAS TRANSFER [J].
BERNARD, TE .
JOURNAL OF APPLIED PHYSIOLOGY, 1977, 43 (02) :375-378
[3]   NEW TECHNIQUE FOR RECORDING RESPIRATORY TRANSIENTS AT START OF EXERCISE [J].
DAVIES, EE ;
HAHN, HL ;
SPIRO, SG ;
EDWARDS, RHT .
RESPIRATION PHYSIOLOGY, 1974, 20 (01) :69-79
[4]  
GEISLER FH, 1978, J TRAUMA, V18
[5]  
GUY HJ, 1976, AM REV RESPIR DIS, V113, P737
[6]  
LINNARSSON D, 1974, SCAND J CLIN LAB INV, V34, P219
[7]  
OSBORN JJ, 1968, SURGERY, V64, P1057
[8]  
OSBORN JJ, 1969, COMPUTERS BIOMEDICAL, V3, P207
[9]  
PEARCE DH, 1977, J APPL PHYSIOL, V42, P968, DOI 10.1152/jappl.1977.42.6.968
[10]   THEORY AND CLINICAL APPLICATION OF A DIGITAL NITROGEN WASHOUT COMPUTER [J].
SHINOZAKI, T ;
ABAJIAN, JC ;
TABAKIN, BS ;
HANSON, JS .
JOURNAL OF APPLIED PHYSIOLOGY, 1966, 21 (01) :202-+