X-RAY FILTER ASSEMBLY FOR FLUORESCENCE MEASUREMENTS OF X-RAY ABSORPTION FINE-STRUCTURE

被引:285
作者
STERN, EA
HEALD, SM
机构
[1] Physics Department, University of Washington, Seattle
基金
美国国家科学基金会;
关键词
D O I
10.1063/1.1135763
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Fluorescence detection, in principle, permits the detection of the extended x-ray absorption fine structure (EXAFS) of more dilute atoms than can be obtained in absorption. To take advantage of this it is necessary, in practice, to eliminate the background that normally accompanies the fluorescence signal. We describe an x-ray filter assembly that accomplishes this purpose. The unique characteristic of the assembly is a slit system that minimizes the fluorescence background from the filter. The theory of the slit assembly is presented and is found to agree with measurements made on the Fe EXAFS of a dilute sample. The filter assembly has a better effective counting rate in this case than that of a crystal monochromator design.
引用
收藏
页码:1579 / 1582
页数:4
相关论文
共 4 条
  • [1] GOULDING FS, 1978, MAY WORKSH XRAY INST
  • [2] HASTINGS JB, 1978, 5TH ANN SSRL US GROU
  • [3] FLUORESCENCE DETECTION OF EXAFS - SENSITIVITY ENHANCEMENT FOR DILUTE SPECIES AND THIN-FILMS
    JAKLEVIC, J
    KIRBY, JA
    KLEIN, MP
    ROBERTSON, AS
    BROWN, GS
    EISENBERGER, P
    [J]. SOLID STATE COMMUNICATIONS, 1977, 23 (09) : 679 - 682
  • [4] TAYLOR A, 1961, XRAY METALLOGRAPHY, P28